Reference
2-12 AFG3000 Series Arbitrary/Function Generators Reference Manual
Output Menu Table 2-9 shows the Output Menu. To access the Output menu, push the
front-panel Top Menu button, and then push the bottom bezel button.
Table2-9: Output Menu
First Level Second Level Description
Load Impedance 50 ΩOption buttons. Set the load impedance to 50 Ω.
Load1
1. Range: 1 Ω to 10 kΩ.
When dBm is currently selected for output amplitude units, the amplitude units setting
is automatically changed to Vp-p if you select high impedance.
Option buttons & Numeric input.
You can set the load impedance to any value
from 1 Ω to 10 kΩ.
High Z Option buttons. Select high impedance.
Invert
Off On
Selects Invert On to invert a waveform.
Limit High Limit Numeric input.
Specifies the higher and lower amplitude limit.
See Level Meter on page1-12.
Low Limit
Noise Noise Add
Off On
Selects Noise Add On to add the internal noise
to a waveform.
Noise Level2
2. Range: 0% to 50% of amplitude setting (Vp-p) of signal waveform, resolution 1%
The maximum value of noise range is limited by waveform and amplitude range.
Sets the noise level.
External Add3
Off On
3. AFG3100 and AFG3200 series only.
Selects External Add On to add an external
signal to the CH 1 output signal.