Measuring Waveforms

CSA8000B & TDS8000B User Manual 3-95

Overview Related control elements and resourcesTo perform a compensation (cont.)
Save the
compensation
7. In the Select Action fields, select Save.
8. Click the Execute button to save the new compensation
results. The new compensation results will be lost when
the instrument is powered down if they are not saved.
The Storage destination for the compensation results is
limited to the User area. The Factory settings cannot be
overwritten.
Recalling a
compensation
9. In the Select Action fields, select Recall.
10. In the Storage field, select the compensation to recall,
UserorFactory.
User recalls the last saved user compensation. Factory
recalls the compensation established at the factory.
Note. Before proceeding, make sure you want to rewrite
the compensation you just saved.
11. Click the Execute button to recall the compensation
results.
Note. The Factory compensation should only be recalled
to bring the instrument to a known state. Recalling the
Factory compensation does not guarantee measurement
accuracy.
End of Procedure