Appendix B: Automatic Measurements Reference

B-6 CSA8000B & TDS8000B User Manual

Table B- 1: Pulse Measurements Amplitude (cont.)
Name Definition
+Overshoot The ratio of the maximum peak to the signal amplitude over the measurement region,
expressed as a percentage.
Where:
HMax is the signal maximum
HHigh and Low are the 100% and 0% reference levels
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement.
+Overshoot =100 ×(Max High)
(High Low)
- Overshoot The ratio of the minimum waveform value to the signal amplitude over the measurement region,
expressed as a percentage.
Where:
HMin is the signal minimum
HHigh and Low are the 100% and 0% signal reference levels.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
For best results with this measurement, optimize the vertical resolution before taking this
measurement.
Overshoot =100 ×(Low Min)
(High Low)