Using Masks, Histograms, and Waveform Databases

CSA8000B & TDS8000B User Manual 3- 147

Overview Related control elements & resourcesTo mask test a waveform (cont.)
Autoset the wave-
form to mask
8. Click the Autoset button to perform a manual autoset on
the mask-source waveform.
Tip. You can choose to autoset the mask-source
waveform to the mask anytime you select a new mask
standard; just check Automatic option under Autoset.
9. Select the HiLow Method used to determine the High
and Low values when aligning the input signal to the
masks.
Mean sets the Mask Autoset to use the mean value of
the High level (topline) and Low level (baseline), taken
within the fixed eye aperture (center 20% of the eye), to
align the input signal to the NRZ mask.
Mode sets the Mask Autoset to use the High level
(topline) and Low level (baseline), taken across one unit
interval of the eye diagram, to align the input signal to
the NRZ mask.
SetStopAction&
start testing
10. From the application menu bar, select Setup, and then
select Acquisition.
11. In the Acq Setup dialog box (see right), check the
Condition option under Stop After.
12. In the Condition pulldown list, select a mask-related
criteria, such as Mask Total Hits and set a count, such
as 1, in the count box.
These settings will stop acquisition when mask
violations satisfy the criteria you set here. See below.
13. Push the RUN/STOP front-panel button to restart
acquisition, if stopped.