Incoming Inspection
1-22 CSA8000B & TDS8000B User Manual
STOP. These procedures verify functions; that is, they verify that the instrument
features operate. They do not verify that they operate within limits; therefore, do
not interpret any quantities cited (such as “about five horizontal divisions”) as
limits.
STOP. DO NOT make changes to the front-panel settings that are not called out
in the procedures. Each verification procedure will require you to set the
instrument to default settings before verifying functions. If you make changes to
these settings, other than those called out in the procedure, you may obtain
invalid results. In this case, go back to step 1 and repeat the procedure.
Install the test hookup and preset the instrument controls:
Equipment
required
One SMA cable, such as Tektronix part number 174-1427-00.
Prerequisites At least one electrical (80E00 series) sampling module must be
installed as outlined in its user manual.
1. Initialize the instrument: Push the front-panel DEFAULT SETUP button,
and click Yes in the confirmation dialog box.
2. Set the Trigger System: In the UI application toolbar, select Internal Clock
from the Trig list box as shown below.
3. Hook up the signal source: Connect the SMA cable from the DC CALIBRA-
TION output to the channel input that you want to test as shown in
Figure 1--7.