Appendix A: Specifications

A-6 CSA8000B & TDS8000B User Manual

Table A- 3: System - Trigger (cont.)
Description Characteristics
Internal clock trigger
rates
Rate selectable at 25, 50, 100, and 200 kHz internally and is provided
to the trigger, to the TDR stimulus drives in small sampling module
interfaces, and to the Internal Clock Out connector on the front panel.
1The input resistance at the external direct trigger input and the maximum input
voltage.
2Maximum signal input for maintaining calibrated time base operation.
3Section 4.10.2 in IEEE standard number 1057. The minimum signal levels required
for stable edge triggering of an acquisition.

Table A- 4: System - Environmental

Description Characteristics
Dynamics Random vibration (operating):
0.22 g rms, from 5 to 500 Hz, 10 minutes each axis, (3 axis,
30 minutes total).
Random vibration (nonoperating):
2.28 g rms, from 5 to 500 Hz, 10 minutes each axis, (3 axis,
30 minutes total) non-operating.
Atmospherics Temperature:
Operating:
10 °Cto+40°C
0°Cto+35°C for 80E0X modules on Tektronix part number
012-1569-00 2-meter extender
Nonoperating:
-- 2 2 °Cto+60°C
Relative humidity:
Operating: 20% to 80%, with a maximum wet bulb temperature
of 29 °C at or below +40 °C (upper limits derates to 45% relative
humidity at +40 °C, non-condensing)
Nonoperating (no floppy disk in floppy drive): 5% to 90%, with a
maximum wet bulb temperature of 29 °C at or below +60 °C (upper
limits derates to 20% relative humidity at +60 °C, non-condensing)
Altitude:
Operating: 3,048 m (10,000 ft.)
Nonoperating: 12,190 m (40,000 ft.)
Electrostatic dis-
charge susceptibility
Up to 8 kV with no change to control settings, or impairment of normal
operation
Up to 15 kV with no damage that prevents recovery of normal operation