Acquiring Waveforms

3-12 CSA8000B & TDS8000B User Manual

Overview Control elements and resourcesTo autoset (cont.)
Define 4. Click Define Autoset in the Utilities menu to display
the Autoset properties dialog box. To change the autoset
criteria, select from:
HEdge to setup the default autoset for instrument to
acquire the waveform data such that the center
20% of the record contains a rising edge.
HPeriod to setup the default autoset for instrument
to acquire the waveform data such that the record
contains 2 or 3 periods.
HNRZ Eye to setup the default autoset for instru-
ment to acquire the waveform data as follows:
Hone bit (two eye crossings) is displayed over
about 7.5 horizontal divisions, centered on the
screen.
Hthe high/low values are displayed over about 6
vertical divisions, also centered on screen.
HRZ Eye Pattern to setup the default autoset for
instrument to acquire the waveform data as follows:
Hthree rise/fall edges are displayed over the
center 6 horizontal divisions, with the first
rising edge placed near the 20% horizontal
location (second division).
Hamplitude (the high/low values) is displayed
over the center 5 vertical divisions.
Click OK to set Autoset to use the current criteria. To
execute, push the Autoset button.
For More
Information
5. For more information on the controls described in this
procedure, push/click the HELP button in any dialog
box or select Help Contents and Index in the Help
menu.
End of Procedure