Appendix A: Specifications

A-2 CSA8000B & TDS8000B User Manual

Table A- 1: System - Signal acquisition (cont.)

Description Characteristics
Small Sampling Mod-
ule Interface
Tekprobe-Sampling Level 3. Hot switching is not permitted on this
interface.
Large Sampling Mod-
ule Interface
Tekprobe-Sampling Level 3. Hot switching is not permitted on this
interface.
1Total actively-acquired channels 8.
Table A- 2: System - Timebase
Description Characteristics
Sampling rate DC-200 kHz maximum, dictated by trigger rate and actual holdoff
setting. If trigger rate is less than the maximum, or the requested
holdoff exceeds the minimum, the trigger rate and/or holdoff will dictate
the sampling rate.
Record length120, 50, 100, 250, 500, 1,000, 2,000, or 4,000 samples.
Horizontal scale
range
1 ps/div to 5 ms/div in 1, 2, 5 steps or 1 ps increments. Maximum
record lengths apply at certain ranges (per table, below).
g
Scale set to an integer multiple of: Maximum record length
1 ps/div 1000
2 ps/div 2000
4 ps/div 4000
Horizontal position
range
50 ms maximum.
Horizontal resolution 10 fs minimum
Horizontal position
setting resolution
1 ps minimum
Horizontal modes Two modes, Short Term Optimized and Locked to 10 MHz Reference.
The 10 MHz reference may be internal or external.
nTime internal ac-
curacy, short term
optimized mode2
Strobe placement accuracy for a given horizontal interval and position
on same strobe line per table below. (Contribution from 80E04
sampling module is included in specification.)
p
Range Time Interval Accuracy
20 ps/div 1 ps + 1% of interval
21 ps/div 8 ps + 0.1% of interval