Installation
CSA8000B & TDS8000B User Manual 1-11
Large-module compartments (2)
Small-module compartments (4)
Connect ESD wrist strap here
Figure 1- 1: Compartments for sampling modules
You can install up to two large sampling modules and four small modules for a
total of 10 inputs. Of these 10 inputs, only eight inputs can be active at one time
(see Figure 1--2, top two configurations). Also, note that installing a single large
module in either compartment disables the first small-module compartment (see
note). This configuration (see Figure 1--2, bottom configuration) limits the input
count to seven—one from the large, six from the small compartments.
NOTE. Power is still provided to this small slot, which does allow an 80A01 to
be functional in this slot even when a large module is installed.
Eight channels: Two large modules and
three small modules
CH 1 CH 2
N.A.1CH 6CH 3 CH 8CH 5 CH 7N.A. CH 4
CH 1 CH 2 CH 6CH 3 CH 8CH 5 CH 7
N.A.
CH 4
N.A.
Eight channels: No large and four
small modules
Seven channels: One large module,
installed in either compartment, and
three small modules
CH 1/N.A.
N.A. CH 6CH 3 CH 5 CH 7N.A. CH 4
CH 2/N.A.
CH 8
1Not Available
Figure 1- 2: Maximum inputs in three configurations
Install probes, cables, and other connection accessories to your sampling
modules as appropriate for your application and sampling module. Again,
consult your sampling-module and connection-accessory manuals. Continue with
the next section after installing the sampling modules.