Incoming Inspection
1-18 CSA8000B & TDS8000B User Manual
HOne SMA 10X attenuator, such as Tektronix part number 015-1003-00.
HOne or more (quantity to match number of electrical channels to compen-
sate) 50 terminators, such as Tektronix part number 015-1022-01
HOne 50 terminator cap, such as Tektronix part number 011-0049-02
HOne 80E00-series electrical sampling modules installed as outlined in its
User manual.
HOne 80C00-series optical sampling module installed as outlined in its User
manual (optional; test only if purchased with/for your instrument).
HMouse
HKeyboard

Perform the Diagnostics

The instrument Diagnostics use internal routines to confirm basic functionality
and proper adjustment.
Equipment required None
Prerequisites First, all sampling modules to be diagnosed must be installed as
outlined in their user manuals.
Second, power on the instrument and allow a 20 minute warm-up
before doing this procedure.
1. Set up the instrument: From the application menu bar, select Utilities, and
then select Diagnostics. The Diagnostics dialog box displays. See below.