CSA8000B & TDS8000B User Manual 1-9
Installation
This section covers installation of the instrument, addressing the following
topics:
HCheck the Environment Requirements on page 1--9
HInstall the Sampling Modules on page 1--10
HConnect the Peripherals on page 1--12
HPower On the Instrument on page 1--13
HPowering Off the Instrument on page 1--15
HBrightness and Contrast Adjustment (Gamma) on page 1--15
HBack Up User Files on page 1--15
The basic operating software is already installed on the hard disk. If reinstalla-
tion of software becomes necessary, see the following topic:
HSoftware Installation on page 1--15
Check the Environmental Requirements
Read this section before attempting any installation procedures. This section
describes site considerations, power requirements, and ground connections for
your instrument.
The instrument is designed to operate on a bench or on a cart in the normal
position (on the bottom feet). For proper cooling, at least two inches (5.1 cm) of
clearance is recommended on the rear and sides of the instrument.
You can also operate the instrument while it rests upright on its rear feet. If you
operate the instrument while it is resting on the rear feet, make sure that you
properly route any cables coming out of the rear of the instrument to avoid
damaging them.
CAUTION. Keep the bottom of the instrument clear of obstructions to ensure
proper cooling.