Appendix A: Specifications
CSA8000B & TDS8000B User Manual A-11
Certifications
Table A- 10: Certifications and compliances
Category Standards or description
EC Declaration of Conformity --
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility when configured with
sampling head modules designed for use with this instrument as identified in this manual.
Compliance was demonstrated to the following specifications as listed in the Official Journal of the
European Union:
EN 61326 EMC Requirements for Electrical Equipment for Measurement, Control
and Laboratory use.
Class A Radiated and Conducted Emissions
IEC 1000-4-2 Performance Criterion B1,2
IEC 1000-4-3 Performance Criterion A1
IEC 1000-4-4 Performance Criterion B1
IEC 1000-4-5 Performance Criterion B1
IEC 1000-4-6 Performance Criterion A1
IEC 1000-4-11 Performance Criterion B1
1Performance Criteria C for USB keyboard and mouse. Note that operation of the
USB keyboard and mouse can be restored by unplugging and then
reconnecting the USB connector at the rear panel of the main instrument.
2Horizontal timing susceptibility of the optical sampling modules and their
internal clock recovery trigger signals usually increase the horizontal timing
jitter when external electromagnetic fields are applied. For fields up to 3 V/m,
the increase in the horizontal high-frequency RMS jitter is typically less than
3 ps RMS of jitter, added using the square-root-of-the-sum-of-the-squares
method. An example follows:
If an 80C01-CR operating in clock-recovery trigger mode exhibits 3.5 ps RMS of
edge jitter, with no EMC field applied and for an ideal jitterless input, then for
applied fields up to 3 V/m the edge jitter, degradation would typically result in a
total RMS jitter of:
Jitter ≤3.5ps2+3ps2
Ꭹ=4.61ps
s
EN 61000-3-2 AC Power Harmonic Current Emissions
Radiated emissions may exceed the levels specified in EN 61326 when this instrument is connected
to a test object.
Australia/New Zealand
Declaration of Conformity --
EMC
Complies with EMC Framework per the following standard:
AS/NZS 2064.1/2 Class A Radiated and Conducted Emissions
General EMC To ensure compliance with EMC requirements, only high quality shielded cables having a reliable,
continuous outer shield (braid & foil) with full coverage, low impedance connections to shielded
connector housings at both ends should be connected to this product.
EC Declaration of Conformity --
Low Voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the
European Union:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC