Teledyne 4060 manual Installation, Unpacking the Analyzer, Mounting the Analyzer

Models: 4060

1 79
Download 79 pages 47.82 Kb
Page 27
Image 27
Installation

Series 4060

Installation

 

 

 

Installation

Installation of the Series 4060 Analyzer includes:

1.Unpacking

2.Mounting

3.Gas connections

4.Electrical connections

5.Testing the system.

3.1Unpacking the Analyzer

Although the analyzer is shipped with all the materials you need to install and prepare the system for operation. Carefully unpack the Analyzer and inspect it for damage. Immediately report any damage or shortages to the shipping agent.

3.2 Mounting the Analyzer

The Series 4060 is a general-purpose analyzer and as such is designed with (non-sealed) enclosures. It must be installed in an area where the ambient temperature is not permitted to drop below 32ºF nor rise above 100ºF. In areas outside these temperatures, auxiliary heating/cooling must be supplied. The 4060 enclosure is oil and dust resistant and although it is designed to resist moisture, it should NOT be considered completely watertight. Mounting to walls or racks must be made securely. Avoid locations that are subject to extreme vibration and sway.

Sufficient space must be provided around the analyzer to accommodate the necessary electrical conduit and plumbing connections. The front panel must be allowed to be pulled out for possible service access to all components of the enclosure. Refer to the system/analyzer outline drawings for dimensions.

Note: To completely slide the analyzer out of the enclosure, pull Analyzer out until it stops, then push down on the release

Teledyne Analytical Instruments

27

Page 27
Image 27
Teledyne 4060 manual Installation, Unpacking the Analyzer, Mounting the Analyzer