Introduction | Series 4060 | |
|
|
|
|
|
|
Extensive
Analog outputs for concentration and range identification (0- 1 VDC standard and isolated
Superior Accuracy
1.2Principle of Operation
The analyzer uses a gas separation column to separate the component of interest, like benzene, from the other hydrocarbons in the sample gas. Using a carrier gas and a microprocessor actuated switching valve, the eluted gas is analyzed in Flame Ionization Detector (FID) and the column is back flushed to accept the fresh sample in the next cycle. More information is provided in Section 2.1 Introduction, under Operational Theory. The electrical output of the
1.3 Analyzer Description
The standard analyzer is a rack mountable instrument designed to fit into a standard 19” instrument rack. The front interface panel is mounted on a door which, when opened, allows convenient access to the Analyzer electronics. The entire front panel can slide out of the chassis to provide greater access to the electronics and to the sample system. Gas pressure and flow controls are mounted on the front panel adjacent to the LED and VFD displays and user interface.
At the rear of the instrument are ports for the introduction of air, fuel, zero, span, and sample gas. A single
Figure 1-1 shows the Model 4060 front interface while Figure 1-2 shows the rear panel including the user connections.
16 | Teledyne Analytical Instruments |