Installation | Series 4060 | |
|
|
|
|
|
|
Addenda included with this manual to determine if special conditions apply.
3.3.3.1EFFLUENT
All the gases introduced into the detection cell vent from one fitting at the rear of the analyzer. TAI recommends that the cell be permitted to vent directly to the atmosphere wherever possible.
If a vent line is required, the installation must include a
3.3.3.2SAMPLE BYPASS VENT
The sample bypassed by the
3.3.3.3FUEL AND AIR CONNECTIONS
The fuel used to provide combustion should be hydrogen gas, zero quality with certified THC of less than 0.5 ppm and supplied at a pressure of 40 psig. The compressed air, ultra zero gas quality with THC less than 0.1 ppm should be used and supplied at a pressure of 40 psig. Connect the fuel and air sources to the instrument according to the gas connection diagram included at the back of this manual.
3.3.3.4SAMPLE AND SPAN GAS CONNECTIONS
The sample and span gas connections are made at the rear panel. If the optional autocal module is installed, the sample and span gas will connect to this module. If no autocal module is present, these gases connect to the labeled fittings on the rear panel.
Set the supply pressure for sample and span gas to 20 psig.
3.4 Placing the System in Operation
See Section 4 for information on starting the analyzer for the first time. Makes sure that all electrical connections have been made correctly and all connectors are fully seated. Make sure all gas connections are correct and
40 | Teledyne Analytical Instruments |