Series 4060

Appendix

 

 

 

Appendix

A.1 Specifications and Initial Settings:

Range(s): Refer to Addendum and Testing Results in Appendix B.

Power Requirements: Refer to Addendum and Testing Results in Appendix B.

Signal Output: Two analog channels:

Channel 1: 0-1 VDC and isolated 4-20 mADC

Channel 2: 0-1 VDC and isolated 4-20 mADC

User configurable as: Benzene, other aromatics, Total HC, and Peak traces.

Alarms: Two Concentration Alarm ‘C’ Type relay contacts. User

configurable.

One System Alarm for AC power failure and flame out, fail safe, ‘C’ type relay contacts.

Calibration Contact: Calibration Contact, ‘A’ Type Relay contact for span mode indication.

RS-232 Output: Provided

Ambient Temp: 0 – 40º C. Install in a well ventilated area