4.6.12 Linearization

Operation

Series 4060

 

 

 

 

______________________________________________

-> CH1 Offset

CH1 Gain

CH1 (Wave, Id for Ch2, or Group Output)

CH1 (Select Groups)

CH2 (Wave, Id for Ch2, or Group Output)

CH2 (Select Groups)

______________________________________________

4.6.12 Linearization

Not applicable to the 4060 configuration.

4.6.13 Changing Stream

Affects the relay on the rear panel that controls which gas is allowed to flow to the column and detector.

When the instrument has electronically controlled calibration valves (option), the gas stream can be switched here. You can flow span, sample (or zero gas) without entering the calibration mode in order to check instrument response. During a span calibration, the span stream is activated. After a span calibration, the sample stream is activated regardless of what was set before the span.

4.6.14 Setting up an AUTO-CAL

The Series 4060 Analyzer can be fitted with an external auto calibration module as an option. This feature provides automatic switching of sample and calibration gases and allows the operator to program calibration events to occur automatically

Note: If your instrument is not fitted with an autocal module, the autocal screens will still appear on the display but will be non-functional.

Note: Before setting up an AUTOCAL, be sure you understand