Installation | Series 4060 | |
|
|
|
|
|
|
3.3.2.2FUSE INSTALLATION
The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used.
3.3.2.3
Figure 3-2 shows the pin layout of the Equipment Interface connector. The arrangement is shown as seen when the viewer faces the rear panel of the analyzer. The pin numbers for each input/output function are given where each function is described in the paragraphs below.
Figure 3-2: Equipment Interface Connector Pin Arrangement
3.3.2.4ANALOG OUTPUT
There are four DC output signal
Voltage rises linearly with increasing | |
| concentration, from 0 V at 0 |
| concentration to 1 V at full scale. (Full |
| scale = 100% of programmable range.) |
0.20 V = Low Range | |
| 0.5 V = Medium Range |
| 0.80 V = High Range |
Current increases linearly with | |
| concentration, from 4 mA at 0 |
| concentration to 20 mA at full scale. |
| (Full scale = 100% of programmable |
| range.) |
6.8 mA = Range 1 | |
| 12.0 mA = Range 2 |
| 16.8 mA = Range 3 |
30 | Teledyne Analytical Instruments |