Installation

Series 4060

 

 

 

 

 

 

3.3.2.2FUSE INSTALLATION

The fuse block, at the right of the power cord receptacle, accepts US or European size fuses. A jumper replaces the fuse in whichever fuse receptacle is not used.

3.3.2.350-PIN EQUIPMENT INTERFACE CONNECTOR

Figure 3-2 shows the pin layout of the Equipment Interface connector. The arrangement is shown as seen when the viewer faces the rear panel of the analyzer. The pin numbers for each input/output function are given where each function is described in the paragraphs below.

Figure 3-2: Equipment Interface Connector Pin Arrangement

3.3.2.4ANALOG OUTPUT

There are four DC output signal pins—two pins per output. For polarity, see Table 3-1. The outputs are:

0–1 VDC % of Range:

Voltage rises linearly with increasing

 

concentration, from 0 V at 0

 

concentration to 1 V at full scale. (Full

 

scale = 100% of programmable range.)

0–1 VDC Range ID:

0.20 V = Low Range

 

0.5 V = Medium Range

 

0.80 V = High Range

4–20 mA DC % Range:

Current increases linearly with

 

concentration, from 4 mA at 0

 

concentration to 20 mA at full scale.

 

(Full scale = 100% of programmable

 

range.)

4–20 mA dc Range ID:

6.8 mA = Range 1

 

12.0 mA = Range 2

 

16.8 mA = Range 3

30

Teledyne Analytical Instruments

Page 30
Image 30
Teledyne 4060 manual Equipment Interface Connector Pin Arrangement