Series 4060

Maintenance

 

 

 

the thyristor gating circuit is continuously delivering gate current from pin 4 of A1, thus maintaining constant full wave AC power to the heater.

When the temperature reaches the selected control point, pin 13 voltage is about the same as pin 9 voltage, and proportional control takes over. The rate at which thyristor (triac) Q1 conducts and allows power to be delivered to the heater is determined by the combination of components R2 & R3, R4, C3, R5, and the thermistor resistance at the control temperature. Consequently, the balance point of the bridge formed by this combination of components can be altered by the selection of R3, causing the circuit to seek a temperature at which the thermistor resistance balances the bridge.

Because IC A1 triggers the thyristor at zero-voltage points in the supply voltage cycle, transient load current surges and radio frequency interference (RFI) are substantially reduced. In addition, use of the zero- voltage-switch reduces the rate of change of on-state current (di/dt) in the thyristor.

5.5.3Electrometer-Amplifier PCB

Schematic No. B79159

Assembly Dwg. No. B79153A

The ions formed in the process of burning hydrogen in the presence of hydrocarbon components of the sample gas cause an electrical conduction between two electrodes in the combustion chamber (or detector cell) that is amplified by a high sensitivity and high input impedance electrometer-amplifier circuit. The electrical output of the electrometer-amplifier is directly proportional to the quantity of flame ionizable hydrocarbons present.

The electrometer amplifier PC board is located on the side of sample module, interconnected to the electronics circuitry by means of a single 8 pins cable, so that the ease of replacement of a board is maintained.

The high input impedance requires a shield, or cover, which is removable for access, as well as a shielded input conductor. Interconnection with the collector is made by a coaxial cable.

Although the cable and fittings are intended for coaxial service, the cable is actually being used as a shielded single-conductor connection. The collector cable plugs into a coaxial connector on the electrometer amplifier PC board, which is located at the side of the sample module.

Teledyne Analytical Instruments

73

Page 73
Image 73
Teledyne 4060 manual Electrometer-Amplifier PCB