Replacing the exposure lamps
The exposure system consists of four
Change the lamps at least every 500 hours, when image quality appears to be degrading (and standard maintenance procedures do not improve the image quality), or when requested to do so by an operator message. Always replace both lamps in a housing at the same time, or exposures may appear inconsistent. Always use the recommended lamps, or image quality may be unacceptable.
1.Turn the side panel switch off.
2.Open the right side access door.
3.Remove the imaging guides. (Refer to the section entitled, “Replacing the imaging guides”.)
4.Grasp the tab at the end of the upper lamp housing and slide the housing out of the track.
Tab
Upper lamp housing
WARNING: Make sure the lamps are cool before attempting to remove them from the lamp housing.
|