IBM CF Series manual

Page 37

The number of resolutions entered in the Global Parameters will determine the functionality of this table. For example, the CF 300 with three resolutions will use motor speed values entered in A through C when different resolutions are selected from the Main Menu.

If you press <T> to tune the speed for Resolution Index A, you will receive the prompt, "Actual [*]," where "*" is the value entered the last time this option was used. Enter the actual measured size of a target you have scanned. You may use any scale of measurement you wish as long as the same units are used to measure the actual target as are used in measuring the scanned image of the target. The measurement may be up to five digits. If it is less than five, press <CR> after entering the value. Then, in response to the next prompt, "Scanned [*]," enter the corresponding measurement of the scanned image.

After entering these two values, the calculated percentage error will appear, followed by the prompt, "Update Resolution A Speed [Y/N]?" An error of .04% or less meets specifications. If the error is greater, reply with "Y" to the prompt to automatically update the Resolution Index A speed.

Press <C> if you wish to recalculate the speeds for the other resolutions based on the value for Resolution Index A.

Press <X> to exit and return to the Diagnostic Tests menu.

Image 37
Contents Firmware Rev. S27G and up Scanner Serial Number CF Series Scanner Maintenance ManualTrademark Notice Preface Related Scangraphics DocumentationIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview Multiple Resolution FeaturesUser Interface Critical CalibrationConstruction DiagnosticsFunctional Specifications SpecificationsNominal Scan Speed/Resolution Chart Physical Specifications Input CharacteristicsPower Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Monthly DailyAnnually Page CF Series Scanner Main Menu On-Board Diagnostic SystemPage This page intentionally left blank Global Parameters Menu Set Global ParametersPage Perform System Initialization Parameter BackupThis page intentionally left blank Set Motor Speed Menu Set Motor SpeedPage Splice Count Listing Example from a CF 1000/44 Splice CountsThis page intentionally left blank Focus and Magnification Menu Focus and MagnificationEquipment, Tools & Materials Required This page intentionally left blank Sample Focus Display FocusPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT 17 Merge and Splice Adjustment Menu Merge and Splice AdjustmentPage Page This page intentionally left blank Line Splice Adjustment Procedure for Adjusting Camera Merge Merge Adjustment24 Set Autosize Parameters Menu Autosize ParametersTest Frame Grabber B SET CCB Registers a26 I/O Board Test Menu Test I/O Board C27 I/O Board Sensor Test 28 Page Offsets Menu Offsets DThis page intentionally left blank 29 Camera System Statistics Camera System Diagnostics EPage NE Normalize Exposure NG Normalize Gain Gain MemoryFill Calibration Memory Display Calibration MemoryTake Line Take Qualified Line Video Board Reset Not used 14.11 L0 Lights OffAutotake 14.12 L1 Lights OnCalibration Check AAAA,DDTest Pattern Control Mirror ControlCalibration Memory Test Camera Delay Factory Use onlyCamera Response Curves HistogramExit 31 Resolution Control Display Resolution ControlPage 32 Set Document Size Menu Document SizesCPU RAM RUN Self TestsRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Proceed With Calibration ? Y/N Autocalibration33 Scsi Buffer Parameters Menu Scsi Buffer ParametersThis page intentionally left blank Pixel Correction and Exposure Values Procedure for Issuing the Save CommandSystem Parameters Page Procedure to Remove Either Side Panel Procedure to Remove the Front CoverProcedure to Remove the Back Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Procedure for Replacing White Background Task -- Replacement of White BackgroundProcedure for Removing and Replacing Camera Assembly Task - Camera Assembly Removal and ReplacementTask -- Camera Calibration Procedure for Adjusting Video BoardExposure Calibration Camera Board OffsetGain Calibration +5 Volts Task - Checking and Adjusting Voltages+12 Volts Volts+120 Volts Procedure for Replacing Fans Task - FAN ReplacementTask - Filter Cleaning and Replacement II. Save the Setup Terminal SetupIII. Host Serial Connection Appendix B Replacement Parts List