IBM CF Series manual NE Normalize Exposure

Page 66

5.1.14.2 NE : Normalize Exposure

Before choosing Normalize Exposure, make sure the following steps are completed:

1.Clean both sides of the scan window.

2.Load clean, white calibration document over entire scan window.

3.Scanner and lamps must be power on for at least 10 minutes.

If you enter “NE” for “Normalize Exposure”, the prompt “BASE EXPOSURE TIME [3F00];” will appear. Reply with <CR>. Next, the prompt “TARGET PIXEL VALUE [C0]:” will appear. Again, reply with <CR>. You will then receive the message “Clearing Gain Memory” followed by the prompt “Use Response Peaks ? [Y]”. Press <CR> to take the positive response.

A response curve will briefly appear for each camera. These are provided for informational purposes only. On the X-axis is the pixel number and on the Y-axis is the light response of the pixel (pixel value). The scanner will then begin automatically calibrating all the cameras, one by one. Messages will appear on the screen telling you which camera is being calibrated. The number you see on the left hand side under the camera number is the exposure time which starts with the number entered at the first prompt. The numbers on the right side are calculated average pixel values. Exposure time decreases by 100 (hex) each time until the target pixel value is reached.Pressing <CR> will abort this operation.

Image 66
Contents CF Series Scanner Maintenance Manual Firmware Rev. S27G and up Scanner Serial NumberTrademark Notice Related Scangraphics Documentation PrefaceIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview User Interface FeaturesMultiple Resolution Critical CalibrationDiagnostics ConstructionSpecifications Functional SpecificationsNominal Scan Speed/Resolution Chart Input Characteristics Physical SpecificationsPower Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Daily MonthlyAnnually Page On-Board Diagnostic System CF Series Scanner Main MenuPage This page intentionally left blank Set Global Parameters Global Parameters MenuPage Parameter Backup Perform System InitializationThis page intentionally left blank Set Motor Speed Set Motor Speed MenuPage Splice Counts Splice Count Listing Example from a CF 1000/44This page intentionally left blank Focus and Magnification Focus and Magnification MenuEquipment, Tools & Materials Required This page intentionally left blank Focus Sample Focus DisplayPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT Merge and Splice Adjustment 17 Merge and Splice Adjustment MenuPage Page This page intentionally left blank Line Splice Adjustment Merge Adjustment Procedure for Adjusting Camera MergeAutosize Parameters 24 Set Autosize Parameters MenuSET CCB Registers a Test Frame Grabber BTest I/O Board C 26 I/O Board Test Menu27 I/O Board Sensor Test Offsets D 28 Page Offsets MenuThis page intentionally left blank Camera System Diagnostics E 29 Camera System StatisticsPage NE Normalize Exposure Gain Memory NG Normalize GainDisplay Calibration Memory Fill Calibration MemoryTake Line Autotake Video Board Reset Not used 14.11 L0 Lights OffTake Qualified Line 14.12 L1 Lights OnAAAA,DD Calibration CheckMirror Control Test Pattern ControlCamera Response Curves Camera Delay Factory Use onlyCalibration Memory Test HistogramExit Resolution Control 31 Resolution Control DisplayPage Document Sizes 32 Set Document Size MenuRUN Self Tests CPU RAMRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Autocalibration Proceed With Calibration ? Y/NScsi Buffer Parameters 33 Scsi Buffer Parameters MenuThis page intentionally left blank Procedure for Issuing the Save Command Pixel Correction and Exposure ValuesSystem Parameters Page Procedure to Remove the Front Cover Procedure to Remove Either Side PanelProcedure to Remove the Back Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Task -- Replacement of White Background Procedure for Replacing White BackgroundTask - Camera Assembly Removal and Replacement Procedure for Removing and Replacing Camera AssemblyProcedure for Adjusting Video Board Task -- Camera CalibrationCamera Board Offset Exposure CalibrationGain Calibration +12 Volts Task - Checking and Adjusting Voltages+5 Volts Volts+120 Volts Task - FAN Replacement Procedure for Replacing FansTask - Filter Cleaning and Replacement Terminal Setup II. Save the SetupIII. Host Serial Connection Appendix B Replacement Parts List