IBM CF Series manual Camera Board Offset, Exposure Calibration

Page 94

6.6.2 Camera Board Offset

Offset is the adjustment for total darkness as perceived by the cameras.

Equipment, Tools & Materials Required:

Small flat head screwdriver or potentiometer tweaker

Camera Board Offset Adjustment Procedure: ( Repeat for Each Camera )

1.Activate “AUTOTAKE” and enter the number of the camera to be adjusted. See section 5.1.14.9 for details.

2.You should either see all zeroes or some numbers on the screen.

3.If you see all zeroes, turn the potentiometer (R17) on the camera board until you start to see non-zero numbers.

4.As you observe the display, turn the potentiometer just until you see zeroes come up on the screen.

6.6.3Exposure Calibration

Exposure calibration compensates for the sensitivity differences between cameras. The imaging lamps must be on for at least 10 minutes before performing this procedure. See section 5.1.14.2 for details on how to use the software.

1.Thoroughly clean both sides of the scan window, using a glass cleaner.

2.Load a sheet of clean, Uniform white paper that covers the entire length of the scan window into the scanner.

3.Select the “NE : Normalize Exposure” command in the Camera System Diagnostics. See section 5.1.14.2 for details.

4.Press “Return” for all the prompts.

Image 94
Contents CF Series Scanner Maintenance Manual Firmware Rev. S27G and up Scanner Serial NumberTrademark Notice Related Scangraphics Documentation PrefaceIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview User Interface FeaturesMultiple Resolution Critical CalibrationDiagnostics ConstructionFunctional Specifications SpecificationsNominal Scan Speed/Resolution Chart Physical Specifications Input CharacteristicsPower Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Monthly DailyAnnually Page On-Board Diagnostic System CF Series Scanner Main MenuPage This page intentionally left blank Set Global Parameters Global Parameters MenuPage Parameter Backup Perform System InitializationThis page intentionally left blank Set Motor Speed Set Motor Speed MenuPage Splice Counts Splice Count Listing Example from a CF 1000/44This page intentionally left blank Focus and Magnification Menu Focus and MagnificationEquipment, Tools & Materials Required This page intentionally left blank Focus Sample Focus DisplayPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT Merge and Splice Adjustment 17 Merge and Splice Adjustment MenuPage Page This page intentionally left blank Line Splice Adjustment Merge Adjustment Procedure for Adjusting Camera MergeAutosize Parameters 24 Set Autosize Parameters MenuSET CCB Registers a Test Frame Grabber BTest I/O Board C 26 I/O Board Test Menu27 I/O Board Sensor Test Offsets D 28 Page Offsets MenuThis page intentionally left blank Camera System Diagnostics E 29 Camera System StatisticsPage NE Normalize Exposure Gain Memory NG Normalize GainFill Calibration Memory Display Calibration MemoryTake Line Autotake Video Board Reset Not used 14.11 L0 Lights OffTake Qualified Line 14.12 L1 Lights OnAAAA,DD Calibration CheckMirror Control Test Pattern ControlCamera Response Curves Camera Delay Factory Use onlyCalibration Memory Test HistogramExit Resolution Control 31 Resolution Control DisplayPage Document Sizes 32 Set Document Size MenuRUN Self Tests CPU RAMRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Autocalibration Proceed With Calibration ? Y/NScsi Buffer Parameters 33 Scsi Buffer Parameters MenuThis page intentionally left blank Pixel Correction and Exposure Values Procedure for Issuing the Save CommandSystem Parameters Page Procedure to Remove Either Side Panel Procedure to Remove the Front CoverProcedure to Remove the Back Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Task -- Replacement of White Background Procedure for Replacing White BackgroundTask - Camera Assembly Removal and Replacement Procedure for Removing and Replacing Camera AssemblyProcedure for Adjusting Video Board Task -- Camera CalibrationCamera Board Offset Exposure CalibrationGain Calibration +12 Volts Task - Checking and Adjusting Voltages+5 Volts Volts+120 Volts Task - FAN Replacement Procedure for Replacing FansTask - Filter Cleaning and Replacement Terminal Setup II. Save the SetupIII. Host Serial Connection Appendix B Replacement Parts List