IBM CF Series manual Procedure for Adjusting Camera Magnification

Page 47

Procedure for Adjusting Camera Magnification:

The first four steps of this procedure -- as well as the required equipment, tools, and materials - - are the same as for adjusting the focus of the camera.

NOTE: In practice, both the focus and magnification of a given camera should be set before moving on to the next camera because of the need for precise placement of the Standard Test Document. Since focus and magnification interact, it is necessary to go back and re-check focus whenever the magnification is changed.

1.Open the top cover and remove the rollers and document plate.

2.Place the Standard Target Document upside down on top of the scan window. The center line of the target (barbell shape) should be directly above the center of the camera being tested.

3.Place the white background strip on top of the Standard Target Document.

4.Place the weight on top of the white background strip.

5.Enter the magnification software by pressing <M>.

6.Use the <C> command to select the appropriate camera.

7.Use the <T> command to set the take mode to "AUTO."

8.Check the result display by the software.

If the message "Can't Find Two Good Bars" is displayed, try adjusting the threshold up or down. If that is unsuccessful, check the placement of the target document. When successful, a distance is reported, and you can go on to the next step.

If a distance is reported, the check is successful, and you can go on to the next step.

9.The magnification software locks its focus on the two rows of parallel lines on the target document. It reports the distance between the leftmost and rightmost lines seen by the camera. Since the lines within each of the two rows are each 50 pixels apart, the distance reported must be a multiple of 50 when the magnification is correct.

Image 47
Contents Firmware Rev. S27G and up Scanner Serial Number CF Series Scanner Maintenance ManualTrademark Notice Preface Related Scangraphics DocumentationIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview Critical Calibration FeaturesMultiple Resolution User InterfaceConstruction DiagnosticsNominal Scan Speed/Resolution Chart SpecificationsFunctional Specifications Power Specifications Input CharacteristicsPhysical Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Annually DailyMonthly Page CF Series Scanner Main Menu On-Board Diagnostic SystemPage This page intentionally left blank Global Parameters Menu Set Global ParametersPage Perform System Initialization Parameter BackupThis page intentionally left blank Set Motor Speed Menu Set Motor SpeedPage Splice Count Listing Example from a CF 1000/44 Splice CountsThis page intentionally left blank Equipment, Tools & Materials Required Focus and MagnificationFocus and Magnification Menu This page intentionally left blank Sample Focus Display FocusPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT 17 Merge and Splice Adjustment Menu Merge and Splice AdjustmentPage Page This page intentionally left blank Line Splice Adjustment Procedure for Adjusting Camera Merge Merge Adjustment24 Set Autosize Parameters Menu Autosize ParametersTest Frame Grabber B SET CCB Registers a26 I/O Board Test Menu Test I/O Board C27 I/O Board Sensor Test 28 Page Offsets Menu Offsets DThis page intentionally left blank 29 Camera System Statistics Camera System Diagnostics EPage NE Normalize Exposure NG Normalize Gain Gain MemoryTake Line Display Calibration MemoryFill Calibration Memory 14.12 L1 Lights On Video Board Reset Not used 14.11 L0 Lights OffTake Qualified Line AutotakeCalibration Check AAAA,DDTest Pattern Control Mirror ControlHistogram Camera Delay Factory Use onlyCalibration Memory Test Camera Response CurvesExit 31 Resolution Control Display Resolution ControlPage 32 Set Document Size Menu Document SizesCPU RAM RUN Self TestsRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Proceed With Calibration ? Y/N Autocalibration33 Scsi Buffer Parameters Menu Scsi Buffer ParametersThis page intentionally left blank System Parameters Procedure for Issuing the Save CommandPixel Correction and Exposure Values Page Procedure to Remove the Back Panel Procedure to Remove the Front CoverProcedure to Remove Either Side Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Procedure for Replacing White Background Task -- Replacement of White BackgroundProcedure for Removing and Replacing Camera Assembly Task - Camera Assembly Removal and ReplacementTask -- Camera Calibration Procedure for Adjusting Video BoardExposure Calibration Camera Board OffsetGain Calibration Volts Task - Checking and Adjusting Voltages+5 Volts +12 Volts+120 Volts Procedure for Replacing Fans Task - FAN ReplacementTask - Filter Cleaning and Replacement II. Save the Setup Terminal SetupIII. Host Serial Connection Appendix B Replacement Parts List