IBM CF Series manual Merge Adjustment, Procedure for Adjusting Camera Merge

Page 57

5.1.8.1.2Adjusting Line Splice Electronically Procedure for Electronic Line Splice Adjustment:

1.Access the Diagnostic Tests Software and select “ Splice and Merge Adjustment.” Answer “yes” to the prompt, “ Use Electronic Splice.”

2.Scan a test document and adjust each boundary.

3.Save the changes to non-volatile RAM (NVRAM). Refer to Section 5.2 for details about the procedure for issuing the SAVE command.

5.1.8.2 Merge Adjustment

NOTE: Cameras must be properly spliced before they can be successfully merged.

Camera merging is the joining of data between adjacent cameras so that data is neither duplicated nor lost between them. The system makes merge adjustment electronically and has a design overlap of 64 to 128 pixels depending on the model of a scanner. To adjust merging, you can delete up to 256 pixels from each camera’s data stream. Each merge adjustment is independent so that adjusting the merge between one pair of cameras does not affect the merge between any other pair of cameras.

Merge adjustment can be made with any high contrast document containing 45 degree lines with clean line edges. (Standard Test Document-- Target area for merging). Any breaks appearing on the monitor will signal a camera merge problem. The standard test document contains a section of lines suitable for testing camera merge. The target area (of the Standard Test Document) for merging is shown in Figure 5-21.

Procedure for Adjusting Camera Merge:

1.Select “Merge and Splice Adjustment” from the Diagnostic Tests Menu.

2.Scan the document and view each boundary. The merge problem will show at the center of the display monitor as a broken line.

3.Adjust the merge using the “ < “ or “ > “ keys until the broken lines is stitched together. Refer to figure 5-19 to see an example of incorrect merge that also shows whether an increase or decrease in value is required to correct the problem.

4.Save the merge adjustment to the non-volatile memory. Refer to section 5.2 for details about the procedure for issuing the SAVE command.

Image 57
Contents Firmware Rev. S27G and up Scanner Serial Number CF Series Scanner Maintenance ManualTrademark Notice Preface Related Scangraphics DocumentationIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview Multiple Resolution FeaturesUser Interface Critical CalibrationConstruction DiagnosticsSpecifications Functional SpecificationsNominal Scan Speed/Resolution Chart Input Characteristics Physical SpecificationsPower Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Daily MonthlyAnnually Page CF Series Scanner Main Menu On-Board Diagnostic SystemPage This page intentionally left blank Global Parameters Menu Set Global ParametersPage Perform System Initialization Parameter BackupThis page intentionally left blank Set Motor Speed Menu Set Motor SpeedPage Splice Count Listing Example from a CF 1000/44 Splice CountsThis page intentionally left blank Focus and Magnification Focus and Magnification MenuEquipment, Tools & Materials Required This page intentionally left blank Sample Focus Display FocusPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT 17 Merge and Splice Adjustment Menu Merge and Splice AdjustmentPage Page This page intentionally left blank Line Splice Adjustment Procedure for Adjusting Camera Merge Merge Adjustment24 Set Autosize Parameters Menu Autosize ParametersTest Frame Grabber B SET CCB Registers a26 I/O Board Test Menu Test I/O Board C27 I/O Board Sensor Test 28 Page Offsets Menu Offsets DThis page intentionally left blank 29 Camera System Statistics Camera System Diagnostics EPage NE Normalize Exposure NG Normalize Gain Gain MemoryDisplay Calibration Memory Fill Calibration MemoryTake Line Take Qualified Line Video Board Reset Not used 14.11 L0 Lights OffAutotake 14.12 L1 Lights OnCalibration Check AAAA,DDTest Pattern Control Mirror ControlCalibration Memory Test Camera Delay Factory Use onlyCamera Response Curves HistogramExit 31 Resolution Control Display Resolution ControlPage 32 Set Document Size Menu Document SizesCPU RAM RUN Self TestsRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Proceed With Calibration ? Y/N Autocalibration33 Scsi Buffer Parameters Menu Scsi Buffer ParametersThis page intentionally left blank Procedure for Issuing the Save Command Pixel Correction and Exposure ValuesSystem Parameters Page Procedure to Remove the Front Cover Procedure to Remove Either Side PanelProcedure to Remove the Back Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Procedure for Replacing White Background Task -- Replacement of White BackgroundProcedure for Removing and Replacing Camera Assembly Task - Camera Assembly Removal and ReplacementTask -- Camera Calibration Procedure for Adjusting Video BoardExposure Calibration Camera Board OffsetGain Calibration +5 Volts Task - Checking and Adjusting Voltages+12 Volts Volts+120 Volts Procedure for Replacing Fans Task - FAN ReplacementTask - Filter Cleaning and Replacement II. Save the Setup Terminal SetupIII. Host Serial Connection Appendix B Replacement Parts List