IBM CF Series manual

Page 44

1.Open the scanner top cover and remove the rollers and document plate.

2.Place the Standard Target Document upside down on top of the scan window. The center line of the target (barbell shape) should be directly above the center of the camera being tested.

3.Place the white background strip on top of the Standard Target Document.

4.Place the weight on top of the white background strip.

5.Enter the focus software by pressing <F>.

6.Use the <C> command to select the appropriate camera.

7.Use the <T> command to set the take mode to "AUTO."

8.Use the <P> command to set the display at pixel 2200.

9.Use the <S> command to search for the target center line.

10.The <S> command will search forward from pixel 2200, looking for asignificant event. If the target is properly centered, the event will be the center line and its position will be near pixel 2650. (The center of a 5312 pixel camera is at 2656). The wave form displayed should be similar to the one illustrated in Figure 5-15.

11.Move the target to the left or right until the position is as close to 2650 as possible. 12.Once you have the wave form on the screen, loosen the screws locking the adjustment of

the lens and then adjust the focus by moving the lens and collar up or down. Your goals is to get the sharpest rising and falling edges possible on the displayed wave form. Changing the focus does affect the magnification.

Image 44
Contents CF Series Scanner Maintenance Manual Firmware Rev. S27G and up Scanner Serial NumberTrademark Notice Related Scangraphics Documentation PrefaceIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview Features Multiple ResolutionUser Interface Critical CalibrationDiagnostics ConstructionNominal Scan Speed/Resolution Chart SpecificationsFunctional Specifications Power Specifications Input CharacteristicsPhysical Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Annually DailyMonthly Page On-Board Diagnostic System CF Series Scanner Main MenuPage This page intentionally left blank Set Global Parameters Global Parameters MenuPage Parameter Backup Perform System InitializationThis page intentionally left blank Set Motor Speed Set Motor Speed MenuPage Splice Counts Splice Count Listing Example from a CF 1000/44This page intentionally left blank Equipment, Tools & Materials Required Focus and MagnificationFocus and Magnification Menu This page intentionally left blank Focus Sample Focus DisplayPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT Merge and Splice Adjustment 17 Merge and Splice Adjustment MenuPage Page This page intentionally left blank Line Splice Adjustment Merge Adjustment Procedure for Adjusting Camera MergeAutosize Parameters 24 Set Autosize Parameters MenuSET CCB Registers a Test Frame Grabber BTest I/O Board C 26 I/O Board Test Menu27 I/O Board Sensor Test Offsets D 28 Page Offsets MenuThis page intentionally left blank Camera System Diagnostics E 29 Camera System StatisticsPage NE Normalize Exposure Gain Memory NG Normalize GainTake Line Display Calibration MemoryFill Calibration Memory Video Board Reset Not used 14.11 L0 Lights Off Take Qualified LineAutotake 14.12 L1 Lights OnAAAA,DD Calibration CheckMirror Control Test Pattern ControlCamera Delay Factory Use only Calibration Memory TestCamera Response Curves HistogramExit Resolution Control 31 Resolution Control DisplayPage Document Sizes 32 Set Document Size MenuRUN Self Tests CPU RAMRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Autocalibration Proceed With Calibration ? Y/NScsi Buffer Parameters 33 Scsi Buffer Parameters MenuThis page intentionally left blank System Parameters Procedure for Issuing the Save CommandPixel Correction and Exposure Values Page Procedure to Remove the Back Panel Procedure to Remove the Front CoverProcedure to Remove Either Side Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Task -- Replacement of White Background Procedure for Replacing White BackgroundTask - Camera Assembly Removal and Replacement Procedure for Removing and Replacing Camera AssemblyProcedure for Adjusting Video Board Task -- Camera CalibrationCamera Board Offset Exposure CalibrationGain Calibration Task - Checking and Adjusting Voltages +5 Volts+12 Volts Volts+120 Volts Task - FAN Replacement Procedure for Replacing FansTask - Filter Cleaning and Replacement Terminal Setup II. Save the SetupIII. Host Serial Connection Appendix B Replacement Parts List