6.7 TASK - - CHECKING AND ADJUSTING VOLTAGES

Refer to the following tables for location and adjustment points of the various power supplies in the system.

____________________________________________________________________

NOTE: The following conventions apply when interpreting these power supply voltage tables:

TB1-6

=

Terminal Block #1, Position #6

PS=

Power Supply

 

J22-1

=

Connector

#22 pin 1 on a PC board

From pin 4 to pin 1 means the positive lead of the meter connects to pin 4 and the negative lead to pin 1 (ground)

____________________________________________________________________

Power usage:

Multibus Card Cage: +5V, +12V, -12V

Video Camera System: +5V, +12V, -12V

Stepper Motor Drive: +5V, +46V (nominal)

• Lamp Power:

+120V (nominal)

+5 Volts:

Check voltages at terminal block TB1 on power supply assembly.

From

To

Reading

Adjust_____

TB1-1

TB2

+5.08 to +5.10 VDC

VR1 on PS1

 

Note: TB1-1 through TB1-4 are bussed together.

 

+12 Volts:

Check voltages at terminal block TB1 on power supply assembly.

From

To

Reading

Adjust________

TB1-5

TB2

+12.04 to +12.06 VDC

+V ADJ on PS2

 

Note: TB1-5 through TB1-7 are bussed together.

 

-12 Volts:

Check voltages at terminal block TB1 on power supply assembly.

From

To

Reading

Adjust_______

 

TB1-8

TB2

-12.04 to -12.06 VDC

-V ADJ on PS2

 

Note: TB1-8 through TB1-10 are bussed together.

 

 

+46 Volts:

Image 96
Contents CF Series Scanner Maintenance Manual Firmware Rev. S27G and up Scanner Serial NumberTrademark Notice Related Scangraphics Documentation PrefaceIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview Features Multiple ResolutionUser Interface Critical CalibrationDiagnostics ConstructionSpecifications Functional SpecificationsNominal Scan Speed/Resolution Chart Input Characteristics Physical SpecificationsPower Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Daily MonthlyAnnually Page On-Board Diagnostic System CF Series Scanner Main MenuPage This page intentionally left blank Set Global Parameters Global Parameters MenuPage Parameter Backup Perform System InitializationThis page intentionally left blank Set Motor Speed Set Motor Speed MenuPage Splice Counts Splice Count Listing Example from a CF 1000/44This page intentionally left blank Focus and Magnification Focus and Magnification MenuEquipment, Tools & Materials Required This page intentionally left blank Focus Sample Focus DisplayPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT Merge and Splice Adjustment 17 Merge and Splice Adjustment MenuPage Page This page intentionally left blank Line Splice Adjustment Merge Adjustment Procedure for Adjusting Camera MergeAutosize Parameters 24 Set Autosize Parameters MenuSET CCB Registers a Test Frame Grabber BTest I/O Board C 26 I/O Board Test Menu27 I/O Board Sensor Test Offsets D 28 Page Offsets MenuThis page intentionally left blank Camera System Diagnostics E 29 Camera System StatisticsPage NE Normalize Exposure Gain Memory NG Normalize GainDisplay Calibration Memory Fill Calibration MemoryTake Line Video Board Reset Not used 14.11 L0 Lights Off Take Qualified LineAutotake 14.12 L1 Lights OnAAAA,DD Calibration CheckMirror Control Test Pattern ControlCamera Delay Factory Use only Calibration Memory TestCamera Response Curves HistogramExit Resolution Control 31 Resolution Control DisplayPage Document Sizes 32 Set Document Size MenuRUN Self Tests CPU RAMRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Autocalibration Proceed With Calibration ? Y/NScsi Buffer Parameters 33 Scsi Buffer Parameters MenuThis page intentionally left blank Procedure for Issuing the Save Command Pixel Correction and Exposure ValuesSystem Parameters Page Procedure to Remove the Front Cover Procedure to Remove Either Side PanelProcedure to Remove the Back Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Task -- Replacement of White Background Procedure for Replacing White BackgroundTask - Camera Assembly Removal and Replacement Procedure for Removing and Replacing Camera AssemblyProcedure for Adjusting Video Board Task -- Camera CalibrationCamera Board Offset Exposure CalibrationGain Calibration Task - Checking and Adjusting Voltages +5 Volts+12 Volts Volts+120 Volts Task - FAN Replacement Procedure for Replacing FansTask - Filter Cleaning and Replacement Terminal Setup II. Save the SetupIII. Host Serial Connection Appendix B Replacement Parts List