5.2 SAVING PARAMETER CHANGES TO NON-VOLATILE MEMORIES

All the programmable system parameters and calibration values are stored in non-volatile memories. They are stored on two different modules. The pixel gain correction values and camera exposure values are stored on the Video Motherboard and are accessed directly when scanning is active. All other parameters are stored on the Calibration and Line Splice ( CALS ) Board.

5.2.1 PIXEL CORRECTION AND EXPOSURE VALUES

Every time camera gain or exposure calibration is performed, new values will directly replace the old values in the memory and the old values are lost.

5.2.2SYSTEM PARAMETERS

Procedure for Issuing the SAVE Command:

In order to save parameter changes permanently in the non-volatile memory, issue either of the following SAVE commands:

From the Main Menu of the Local Control Terminal, press and hold down the <CTRL> key as you press <S> to save the changes made.

From the scanner’s keypad, press the <OPT> key and then press the down arrow key. On

the left side of the keypad display, “4. Save Parameters “ will appear. On the right side of the display , two messages will alternate : “ A : Proceed “ and “ OPT : Exit “. Press the keypad <A> key to save parameters followed by <OPT> to exit .

In either case, the system will beep to indicate that the command has been executed.

______________________________________________________________________

NOTE: Parameters may be saved only when the memory write protect switch on the CALS Board in set to “ Write Enable “. Since this is the default switch setting, it is not necessary to write-protect the non-volatile memory.

______________________________________________________________________

If the scanner returns the message, “SORRY, WRITE PROTECTED”, perform the following procedure to change the write protect status of the CALS Board first:

1. Open the front cover of the scanner and the card-cage door.

Image 86
Contents CF Series Scanner Maintenance Manual Firmware Rev. S27G and up Scanner Serial NumberTrademark Notice Related Scangraphics Documentation PrefaceIf You Have Questions Table of Contents Connections ON-BOARD Diagnostic System Diagnostics Tests Saving Parameter Changes to NON-VOLATILE Memories  System Overview User Interface FeaturesMultiple Resolution Critical CalibrationDiagnostics ConstructionNominal Scan Speed/Resolution Chart SpecificationsFunctional Specifications Power Specifications Input CharacteristicsPhysical Specifications Environmental Specifications This page intentionally left blank Installation Connections POWER-UP DE-INSTALLATION System Description Major Systems of Scanner Model # of Cameras Page Page This page intentionally left blank Annually DailyMonthly Page On-Board Diagnostic System CF Series Scanner Main MenuPage This page intentionally left blank Set Global Parameters Global Parameters MenuPage Parameter Backup Perform System InitializationThis page intentionally left blank Set Motor Speed Set Motor Speed MenuPage Splice Counts Splice Count Listing Example from a CF 1000/44This page intentionally left blank Equipment, Tools & Materials Required Focus and MagnificationFocus and Magnification Menu This page intentionally left blank Focus Sample Focus DisplayPage Page This page intentionally left blank Magnification Procedure for Adjusting Camera Magnification Magnification Values for CF300 This page intentionally left blank IP LUT For the Image Processor Board Example of an Image Processor Board Look-Up Table IP LUT Merge and Splice Adjustment 17 Merge and Splice Adjustment MenuPage Page This page intentionally left blank Line Splice Adjustment Merge Adjustment Procedure for Adjusting Camera MergeAutosize Parameters 24 Set Autosize Parameters MenuSET CCB Registers a Test Frame Grabber BTest I/O Board C 26 I/O Board Test Menu27 I/O Board Sensor Test Offsets D 28 Page Offsets MenuThis page intentionally left blank Camera System Diagnostics E 29 Camera System StatisticsPage NE Normalize Exposure Gain Memory NG Normalize GainTake Line Display Calibration MemoryFill Calibration Memory Autotake Video Board Reset Not used 14.11 L0 Lights OffTake Qualified Line 14.12 L1 Lights OnAAAA,DD Calibration CheckMirror Control Test Pattern ControlCamera Response Curves Camera Delay Factory Use onlyCalibration Memory Test HistogramExit Resolution Control 31 Resolution Control DisplayPage Document Sizes 32 Set Document Size MenuRUN Self Tests CPU RAMRestore Nvram This page intentionally left blank Binary Filter Parameters Prompt Parameter Set A,B,C,D will appear IP Averager LUT Autocalibration Proceed With Calibration ? Y/NScsi Buffer Parameters 33 Scsi Buffer Parameters MenuThis page intentionally left blank System Parameters Procedure for Issuing the Save CommandPixel Correction and Exposure Values Page Procedure to Remove the Back Panel Procedure to Remove the Front CoverProcedure to Remove Either Side Panel Procedure for Replacing Belt and Adjusting Tension Procedure for Installing and Replacing Imaging Lamp Task -- Replacement of White Background Procedure for Replacing White BackgroundTask - Camera Assembly Removal and Replacement Procedure for Removing and Replacing Camera AssemblyProcedure for Adjusting Video Board Task -- Camera CalibrationCamera Board Offset Exposure CalibrationGain Calibration +12 Volts Task - Checking and Adjusting Voltages+5 Volts Volts+120 Volts Task - FAN Replacement Procedure for Replacing FansTask - Filter Cleaning and Replacement Terminal Setup II. Save the SetupIII. Host Serial Connection Appendix B Replacement Parts List