Parameter | Result |
|
|
|
|
PCS | Pass if PCS > 75%,RL > 25% and RD < 30% |
| Fail if not satisfied |
|
|
RI/Rd | Pass if Rd/RI < 50% |
| Fail if Rd/RI > 50% |
|
|
SymCont | A if > 70% |
| B if > 55% |
| C if > 40% |
| D if > 20% |
| F if < 20% |
|
|
Edge contrast (min) | A if > 15% |
| F if < 15% |
|
|
Modulation | A if > .70 |
| B if > .60 |
| C if > .50 |
| D if > .40 |
| F if < .40 |
|
|
Defects | A if < .15 |
| B if < .20 |
| C if < ..25 |
| D if < .30 |
| F if > .35 |
|
|
* For UPC, the minimum P.C.S. varies depending on the measured
Errors in Reflectance Parameters
If, after a scan, the abbreviation for Reflectance appears as >REF<, an error exists in one or more of the reflectance parameters. If so, you should use the arrow buttons to review each parameter. If the abbreviation for Reflectance appears as Ref, the symbol has no reflectance errors.
Print Contrast Signal, Light and Dark Reflectance
During a scan, the Quick Check Verifier obtains the maximum light reflectance from spaces (Rmax or RI) and minimum dark reflectance from bars (Rmin or Rd). The print contrast signal, “P.C.S.,” is a traditional measure of print contrast.
PCS=
The traditional parameters are rated OK! or pass if the PCS, RI and Rd meet these criteria. If a value does not meet the criteria, the Ref parameters fail (LO!). If the PCS, RI or Rd values appears on the display or on a printout as “XX” or “—