During Use

 

Discharge static electricity from the weighing material. When

Material case

weighing sample (plastics, insulator, etc.) could have a static

Charged

charge, the weighing value is influenced. Ground the balance, and

sample

Eliminate the static electricity by using an optinal static

 

eliminator, AD-1683.

Grounding

Or try to keep the ambient humidity above 45%RH at the room.

Or use the metal shield case.

 

Or wipe a charged material (plastic sample etc.) with the wet cloth.

 

This balance uses a strong magnet as part of the balance

 

assembly, so please use caution when weighing magnetic

Magnetic

materials such as iron. If there is a problem, use the underhook on

material

the bottom of the balance to suspend the material away from the

 

influence of the magnet.

Draft

Eliminate any temperature difference between the sample and the

20°C 40°C

environment. When a sample is warmer (cooler) than the ambient

temperature, the sample will be lighter (heavier) than the true

Weighing Pan

weight. This error is due to a rising (falling) draft around the sample.

 

Make each weighing gently and quickly to avoid errors due to

Shock

changes in the environmental conditions.

 

Do not drop things upon the weighing pan, or place a sample on the pan that is beyond the balance weighing capacity. Place the sample in the center of the weighing pan.

Do not use a sharp instrument such as a pencil to press the keys.

Use your finger only.

Press the RE-ZEROpossible errors.

key before each weighing to prevent

Calibrate the balance periodically so as to eliminate possible errors.

Take into consideration the affect of air buoyancy on a sample when more accuracy is required.

It is possible to check the reference card for principle operation.

Prevent foreign matter, such as powder, liquid and metal, from invading the area around the weighing pan.

Use the "breeze break" for a precision weighing.

3-3. After Use

Avoid mechanical shock to the balance.

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.

Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is moistened with warm water and a mild detergent.

9

Image 11 Contents
1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 14-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 10-510-6 About This Manual FeaturesIntroduction Compliance with the following Council Directives Compliance To keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Do not allow vibration or drafts to affect Self Check Function with Response AdjustmentBalance during adjustment Terms CalibrationCalibration Group 150 g ~ +0.159 g Automatic self calibration due to changes in temperatureAutomatic Self Calibration About the internal mass Calibration using the internal massOne-Touch Calibration When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output modeCp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableDisplay character set Setting the ID NumberID Number And GLP Report Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hSignature Space, Ascii 20h Calibration report using an external weightCalibration test report using an external weight End block Title block and end blockOperation Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersOption RS-232C TerminalsExternal contact input Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited