A&D MC-10K, MC-30K manual Calibration Group, Terms

Page 20

8. Calibration

8-1. Calibration Group

The balance has the following modes as a calibration group.

Calibration

Automatic self calibration (Calibration due to changes in temperature)

Calibration using the internal mass (One-touch calibration)

Calibration using an external weight

Calibration Test

Calibration test using an external weight (Calibration test does not perform calibration)

Correction of the internal mass value

Terms

The following terms are defined as follows:

Internal mass

= Built-in calibration weight

External weight

= A weight that you have. Referred to as a calibration weight when

 

used for calibration.

Calibration weight

= A weight used for calibration

Target weight

= An external weight used for calibration test

Caution

Calibration adjusts the balance for accurate weighing.

Besides periodic calibration and before each use, perform calibration when:

The balance is installed for the first time.

The balance has been moved.

The ambient environment has changed.

Do not allow vibration or drafts to affect the balance during calibration.

To output the data for GLP using the RS-232C interface, set "GLP output (info)" of "Data output (dout)". Refer to "10. Function Table". Time and date can be added to the GLP report. If the time or date is not correct, adjust them. Refer to the GX-K series instruction manual "10-7. Clock and Calendar Function".

Calibration test is available only when "GLP output (info)" of "Data output ( dout )" is set to "1" or "2",

The calibration and calibration test data can be stored in memory. To store them, set "Data memory (data)" to "3". Refer to the GX-K series instruction manual "12. Data Memory" for details.

The value of the internal mass may change due to aging, corrosion or other damage caused by the operating environment. Check the internal mass periodically. Correct the internal mass value as necessary.

18

Image 20
Contents MC-10K MC-30K 1WMPD4002203AThis Manual and Marks All rights reservedContents 14-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 10-510-6 About This Manual FeaturesIntroduction Compliance Compliance with the following Council DirectivesUnpacking And Installing The Balance To keep waterproof DustproofAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Installing the Balance EarthPrecautions Before UseDuring Use After UsePower Supply Display Symbols And Key Operation Key When pressed and released When pressed and heldKey Operations DisplayWeighing Units Tael HK general, Singapore Selecting a Weighing Unit Mode WeighingBasic Weighing F1l Filter Used DisplayWhen Building into a System Weighing Speed Adjustment Do not allow vibration or drafts to affect Self Check Function with Response AdjustmentBalance during adjustment Terms CalibrationCalibration Group 150 g ~ +0.159 g Automatic self calibration due to changes in temperatureAutomatic Self Calibration About the internal mass Calibration using the internal massOne-Touch Calibration Calibration Using an External Weight When the MC-10K is usedCalibration Test Using an External Weight Print key...... To store the new weight value. Even ifRemove Remove the weight from the pan Step Balance displays Step Confirm that there is nothing on the pan and pressCorrecting the Internal Mass Value Key Function Switch And Initialization Permit or InhibitTo cancel This operation, press Key Initializing the BalanceStep Press Key Step Press Key to displayCheck the settings and condition before changing parameters Function TableStructure and Sequence of the Function Table Display and Operation KeysDetails of the Function Table Data output mode Cp fnc Cp-bBep LO buzzerTimeout AK, Error codeTime/Date output Dout ID number outputID number setting Internal mass correctionF1-b Averaging range for First moving Average F1-tZero upon power-on p-tr Description of the Class Environment, DisplayCondition Cond Stability band width 5t-bDescription of the Item Data Output Mode Key modeStream mode Auto print modes a and BInterval memory mode Description of the Item Data FormatStandard format 5if type DP Dump print format 5if typeKF format MT formatNU numerical format 5if type CSV formatData number DateTime Tare valueNegative error Data Format ExamplesStable OverloadDisplay character set Setting the ID NumberID Number And GLP Report Calibration report using the internal mass GLP ReportSpace, Ascii 20h Terminator, CR , LF or CRSignature Space, Ascii 20h Calibration report using an external weightCalibration test report using an external weight End block Title block and end blockOperation Standard Input And Output Interface RS-232C and External Contact InputSub 25 pin numbers Sub 25 pin assignmentsOption RS-232C TerminalsExternal contact input Settings Not Auto print polarityNecessary Data output pauseUsing Windows Communication Tools Software WinCT Using the WinCT software, the balance can do the followingRsCom RsKeyCommands Command ListAcknowledge Code and Error Codes Control Using CTS and RTS Settings Related to RS-232CCommand Personal computer Time Balance Command Weighing modeExtended Function Page Page Maintenance Treatment of the BalanceTroubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Weighing method16-2 Automatic response adjustment zero error Automatic response adjustment unstable errorInternal error Clock battery errorOther Display Asking for RepairSpecifications External Dimensions Unit mmAD-8922A Remote controller Options and Peripheral InstrumentsAX-MC10K/30KPAN Auto-centering Pan AD-8121B PrinterRechargeable Battery Foot SwitchWeighing Environment Logger Data LoggerTerms/Index TermsMemory function Nothing is on the weighing panKeys and symbols IndexError Mode MT formatF1l FUlPage UK Office German Sales OfficeKorea Limited Instruments India Private Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.