A&D MC-30K, MC-10K manual Using Windows Communication Tools Software WinCT, RsCom, RsKey

Page 47

12-2-2. Connection to a Computer and the Use of WinCT

The balance is of the DCE type (Data Communication Equipment), which can be connected to a personal computer using the RS-232C interface. Before connection, read the personal computer manual thoroughly. Use a standard DCE cable for connection (cable type: straight-through). When the personal computer type is a DOS/V with a 9-pin port, use a straight-through cable with a 25-pin male connector and a 9-pin female connector.

Using Windows Communication Tools Software (WinCT)

When Windows is used as an operating system in a personal computer, the provided WinCT software can be used to transmit the weighing data to the personal computer.

The WinCT software has two communication methods: "RsCom" and "RsKey". Refer to the WinCT instruction manual.

RsCom

RsCom can transmit commands to control the balance.

RsCom can perform bi-directional communication between the balance and a personal computer using the RS-232C interface.

RsCom can display or store the data using a text file format. RsCom can also print the data using a printer connected to the personal computer.

When several ports of a personal computer have balances connected, can communicate with each balance simultaneously.

RsCom can share a personal computer with other application software. RsCom can receive the balance GLP report.

RsKey

RsKey can transmit the weighing data output from the balance directly to other application software such as Microsoft Excel.

RsKey can be used with most application software. RsKey can receive the balance GLP report.

Using the WinCT software, the balance can do the following:

Analyzing the weighing data and the statistics with "RsKey"

The weighing data can be input directly into an Excel worksheet. Then, Excel can analyze the data to obtain total, average, standard deviation, maximum and minimum value, and display them in a graph.

Controlling the balance using commands from a personal computer

By using "RsCom", the personal computer sends commands such as "re-zero" or "send weighing data" to the balance and controls the balance.

Printing the balance GLP report using your printer

The balance GLP report can be printed using a printer connected to the personal computer.

Receiving weighing data at a certain interval

The weighing data can be received at a certain interval and data characteristic with elapsed time can be obtained.

Using the balance memory function

The data can be stored in the balance’s memory. Of the data stored, the weighing data and calibration data can be transmitted to a personal computer at one time.

Using a personal computer as an external indicator

With the "RsKey" test mode function, a personal computer can be used as an external weight indicator for the balance. (To do this, set the balance data output mode to stream mode.)

45

Image 47
Contents 1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 14-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 10-510-6 About This Manual FeaturesIntroduction Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Do not allow vibration or drafts to affect Self Check Function with Response AdjustmentBalance during adjustment Terms CalibrationCalibration Group 150 g ~ +0.159 g Automatic self calibration due to changes in temperatureAutomatic Self Calibration About the internal mass Calibration using the internal massOne-Touch Calibration When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output modeCp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableDisplay character set Setting the ID NumberID Number And GLP Report Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hSignature Space, Ascii 20h Calibration report using an external weightCalibration test report using an external weight End block Title block and end blockOperation Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersOption RS-232C TerminalsExternal contact input Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.