A&D MC-10K, MC-30K manual Commands, Command List

Page 48

13. Commands

13-1. Command List

Note A command has a terminator added, that is specified using "5if Crlf" of the function table, and is sent to the balance.

Commands to query weighing data

C

 

 

 

 

Cancels the S or SIR command.

Q

 

 

 

 

Requests the weighing data immediately.

S

 

 

 

 

Requests the weighing data when stabilized.

SI

 

 

 

 

Requests the weighing data immediately.

SIR

 

 

 

 

Requests the weighing data continuously.

 

 

 

 

 

Commands to control the balance

?CN

 

 

 

 

Requests the upper/lower limit value code number of the selected value.

?HI

 

 

 

 

Requests the upper limit value.

?ID

 

 

 

 

Requests the identification number.

?LO

 

 

 

 

Requests the lower limit value.

?MA

 

 

 

 

Outputs all weighing data in memory.

?MQnnn

 

 

 

Outputs data with the data number nnn. nnn: Three digits

?MX

 

 

 

 

Outputs the number of data in memory (the last data number)

?PN

 

 

 

 

Request the tare number of the selected value.

?PT

 

 

 

 

Request the tare value.

?SN

 

 

 

 

Request the serial number of the balance.

?TN

 

 

 

 

Request the model name of the balance.

?UN

 

 

 

 

Requests the unit mass numbers in memory.

?UW

 

 

 

 

Requests the unit mass value.

CAL

 

 

 

 

Same as the

 

key.

 

 

 

 

CAL

 

 

 

 

 

 

 

CN:mm

 

 

 

Recalls the upper/lower limit value in memory. mm: 01 to 20.

 

∗∗∗∗∗∗

 

~~

 

Sets the upper limit values. ~ is space mark.

HI:

.

g

Example: the upper limit value is 2000.0 g.

 

 

Command: HI:+002000.0~~g

 

 

 

 

 

ID:∗∗∗∗∗∗∗

 

 

Sets identification number.

 

∗∗∗∗∗∗

 

~~

 

Sets the lower limit values. ~ is space mark.

LO:

.

g

Example: the lower limit value is 1000.0 g.

 

 

Command :LO:+001000.0~~g

 

 

 

 

 

MCL

 

 

 

 

Deletes all data in memory.

MD:nnn

 

 

 

Deletes data with the data number nnn. nnn: Three digits.

OFF

 

 

 

 

Turns the display off.

ON

 

 

 

 

Turns the display on.

P

 

 

 

 

Same as the

 

 

 

key

 

 

 

 

ON:OFF

 

 

 

 

 

 

PN:mm

 

 

 

Recalls the tare value in memory. mm: 01 to 20.

PRT

 

 

 

 

Same as the

 

 

key

 

 

 

 

PRINT

 

 

 

 

 

 

PT:∗∗∗∗∗.~~g

Sets the tare value. ~ is space mark.

Example: the tare value is 1000.0 g.

 

 

 

 

 

Command :PT:+001000.0~~g

 

 

 

 

 

46

Image 48
Contents MC-10K MC-30K 1WMPD4002203AThis Manual and Marks All rights reservedContents 10-5 10-614-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Features IntroductionAbout This Manual Compliance Compliance with the following Council DirectivesUnpacking And Installing The Balance To keep waterproof DustproofAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Installing the Balance EarthPrecautions Before UseDuring Use After UsePower Supply Display Symbols And Key Operation Key When pressed and released When pressed and heldKey Operations DisplayWeighing Units Tael HK general, Singapore Selecting a Weighing Unit Mode WeighingBasic Weighing F1l Filter Used DisplayWhen Building into a System Weighing Speed Adjustment Self Check Function with Response Adjustment Balance during adjustmentDo not allow vibration or drafts to affect Calibration Calibration GroupTerms Automatic self calibration due to changes in temperature Automatic Self Calibration150 g ~ +0.159 g Calibration using the internal mass One-Touch CalibrationAbout the internal mass Calibration Using an External Weight When the MC-10K is usedCalibration Test Using an External Weight Print key...... To store the new weight value. Even ifRemove Remove the weight from the pan Step Balance displays Step Confirm that there is nothing on the pan and pressCorrecting the Internal Mass Value Key Function Switch And Initialization Permit or InhibitTo cancel This operation, press Key Initializing the BalanceStep Press Key Step Press Key to displayCheck the settings and condition before changing parameters Function TableStructure and Sequence of the Function Table Display and Operation KeysDetails of the Function Table Data output mode Cp fnc Cp-bBep LO buzzerTimeout AK, Error codeTime/Date output Dout ID number outputID number setting Internal mass correctionF1-b Averaging range for First moving Average F1-tZero upon power-on p-tr Description of the Class Environment, DisplayCondition Cond Stability band width 5t-bDescription of the Item Data Output Mode Key modeStream mode Auto print modes a and BInterval memory mode Description of the Item Data FormatStandard format 5if type DP Dump print format 5if typeKF format MT formatNU numerical format 5if type CSV formatData number DateTime Tare valueNegative error Data Format ExamplesStable OverloadSetting the ID Number ID Number And GLP ReportDisplay character set Calibration report using the internal mass GLP ReportSpace, Ascii 20h Terminator, CR , LF or CRCalibration report using an external weight Calibration test report using an external weightSignature Space, Ascii 20h Title block and end block OperationEnd block Standard Input And Output Interface RS-232C and External Contact InputSub 25 pin numbers Sub 25 pin assignmentsRS-232C Terminals External contact inputOption Settings Not Auto print polarityNecessary Data output pauseUsing Windows Communication Tools Software WinCT Using the WinCT software, the balance can do the followingRsCom RsKeyCommands Command ListAcknowledge Code and Error Codes Control Using CTS and RTS Settings Related to RS-232CCommand Personal computer Time Balance Command Weighing modeExtended Function Page Page Maintenance Treatment of the BalanceTroubleshooting Checking the Balance Performance and EnvironmentChecking that the balance performs properly Weighing method16-2 Automatic response adjustment zero error Automatic response adjustment unstable errorInternal error Clock battery errorOther Display Asking for RepairSpecifications External Dimensions Unit mmAD-8922A Remote controller Options and Peripheral InstrumentsAX-MC10K/30KPAN Auto-centering Pan AD-8121B PrinterRechargeable Battery Foot SwitchWeighing Environment Logger Data LoggerTerms/Index TermsMemory function Nothing is on the weighing panKeys and symbols IndexError Mode MT formatF1l FUlPage UK Office German Sales OfficeKorea Limited Instruments India Private Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.