A&D MC-30K Cp fnc Cp-b, Bep, LO buzzer, OK buzzer, HI buzzer, Cp Hi Upper limit, Cp lo, Cp lmt

Page 31

Class

Item and Parameter

 

Description

Cp fnc

Cp-b

 

 

 

0

OFF

Displays the results on the main

 

Comparator

Main display comparison

 

 

 

1

ON

portion of the display in place of the

 

weight value.

Displayed

bep

 

 

 

0

OFF

Select whether or not to sound the

 

 

only when

LO buzzer

 

1

ON

LO buzzer.

Comparator

 

 

 

 

 

 

 

 

 

bep

 

 

 

0

OFF

Select whether or not to sound the

output

 

 

 

 

OK buzzer

 

 

 

1

ON

OK buzzer.

(GX-04K) is

 

installed

 

 

 

 

 

 

 

 

 

bep

 

 

 

0

OFF

Select whether or not to sound the

 

 

 

 

HI buzzer

 

1

ON

HI buzzer.

Cp Hi

 

 

 

 

 

 

 

 

 

Upper limit

 

 

 

 

 

Refer to the GX-K manual

Displayed when Cp in 0

Cp lo

 

 

 

 

 

"10-8. Comparator Function"

is selected.

Lower limit

 

 

 

 

 

 

 

 

 

Cp ref

 

 

 

 

 

 

 

 

 

Reference value

 

Refer to the GX-K manual

Displayed when Cp in 1

Cp lmt

 

 

 

 

 

"10-8. Comparator Function"

is selected.

Tolerance

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accepts the

 

key only when

 

 

 

 

 

 

0

Key mode

PRINT

 

 

 

 

 

 

the display is stable.

 

 

 

 

 

 

1

Auto print mode A

Outputs data when the display is

 

 

prt

(Reference = zero)

 

 

Data output mode

 

Auto print mode B

stable and conditions of ap-p, ap-b

 

 

2

and the reference value are met.

 

 

 

 

 

 

 

(Reference = last stable value)

With data 0, outputs data

 

 

 

 

 

 

3

Stream mode /

 

 

 

 

 

 

continuously; with data 2, uses

 

 

 

 

 

 

 

Interval memory mode

interval memory.

 

 

 

 

 

 

0

Plus only

Displayed value>Reference

 

 

 

 

 

 

 

 

ap-p

 

 

1

Minus only

Displayed value<Reference

 

 

Auto print polarity

 

 

2

Both

Regardless of displayed value

 

 

 

 

 

 

 

 

 

 

 

 

0

10 digits

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ap-b

Difference between reference value

 

 

1

100 digits

 

 

Auto print difference

and displayed value

 

 

2

1000 digits

 

 

 

 

 

 

 

 

 

dout

 

 

 

 

0

Not used

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

Stores unit mass in counting mode

 

 

 

Data output

 

 

 

 

 

 

 

 

 

data

2

Stores weighing data

Related items:

 

 

Data memory

3

Stores calibration data

prt,int, d-no, 5-td, info

4Stores comparator settings

5Stores tare value

 

 

0

Every measurement

 

 

 

 

 

 

 

 

1

2 seconds

 

 

2

5 seconds

 

int

3

10 seconds

Interval time in the interval memory

4

30 seconds

Interval time

mode when using prt 3,data 2

5

1 minute

 

 

 

6

2 minute

 

 

7

5 minute

 

 

8

10 minute

 

: Factory settings. Digit is a unit of minimum weighing value. 29

Image 31 Contents
1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 10-6 10-514-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Introduction FeaturesAbout This Manual Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Balance during adjustment Self Check Function with Response AdjustmentDo not allow vibration or drafts to affect Calibration Group CalibrationTerms Automatic Self Calibration Automatic self calibration due to changes in temperature150 g ~ +0.159 g One-Touch Calibration Calibration using the internal massAbout the internal mass When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output modeCp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableID Number And GLP Report Setting the ID NumberDisplay character set Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hCalibration test report using an external weight Calibration report using an external weightSignature Space, Ascii 20h Operation Title block and end blockEnd block Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersExternal contact input RS-232C TerminalsOption Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited