7-2. Self Check Function with Response Adjustment
This function automatically updates the response adjustment by analyzing the influence of the environment on the weighing data and also
the internal mass.
Step | 1 | Press and hold the | MODE | key until | RESPONSE | is |
|
| displayed, and then | release | the key. |
|
|
Step | 2 | The balance automatically starts to check the balance | ||||
|
| performance and sets the response characteristic. | ||||
Caution | Do not allow vibration or drafts to affect the | |||||
|
| balance during adjustment. |
Press and hold
Release
CH . . . .
Result
CH
Step 3 After automatic adjustment, the balance displays the updated | 0.000 g | |
response indicator and returns to the weighing mode. | ||
| ||
The response indicator remains displayed for a while. |
| |
Example of display |
|
MID. and
OK
:The example above indicates that the result of the self check is good and MID. is selected as the response rate..
Note
If improper performance is found in the self check, the balance displays CH no .
Contact the local A&D dealer for repair.
If the automatic response adjustment fails, the balance displays CH ng . Check the ambient conditions such as breeze and vibration, also check the weighing pan. Then, perform the adjustment again. To return to the weighing mode, press the CAL key.