A&D MC-30K, MC-10K manual Self Check Function with Response Adjustment, Balance during adjustment

Page 19

7-2. Self Check Function with Response Adjustment

This function automatically updates the response adjustment by analyzing the influence of the environment on the weighing data and also self-checks the balance performance using

the internal mass.

Step

1

Press and hold the

MODE

key until

RESPONSE

is

 

 

displayed, and then

release

the key.

 

 

Step

2

The balance automatically starts to check the balance

 

 

performance and sets the response characteristic.

Caution

Do not allow vibration or drafts to affect the

 

 

balance during adjustment.

Press and hold

Release

CH . . . .

Result

CH

Step 3 After automatic adjustment, the balance displays the updated

0.000 g

response indicator and returns to the weighing mode.

 

The response indicator remains displayed for a while.

 

Example of display

 

MID. and

OK

:The example above indicates that the result of the self check is good and MID. is selected as the response rate..

Note

If improper performance is found in the self check, the balance displays CH no .

Contact the local A&D dealer for repair.

If the automatic response adjustment fails, the balance displays CH ng . Check the ambient conditions such as breeze and vibration, also check the weighing pan. Then, perform the adjustment again. To return to the weighing mode, press the CAL key.

17

Image 19 Contents
1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 10-6 10-514-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Introduction FeaturesAbout This Manual Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Balance during adjustment Self Check Function with Response AdjustmentDo not allow vibration or drafts to affect Calibration Group CalibrationTerms Automatic Self Calibration Automatic self calibration due to changes in temperature150 g ~ +0.159 g One-Touch Calibration Calibration using the internal massAbout the internal mass When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output modeCp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableID Number And GLP Report Setting the ID NumberDisplay character set Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hCalibration test report using an external weight Calibration report using an external weightSignature Space, Ascii 20h Operation Title block and end blockEnd block Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersExternal contact input RS-232C TerminalsOption Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited