A&D MC-30K, MC-10K Description of the Item Data Output Mode, Key mode, Auto print modes a and B

Page 35

10-5. Description of the Item "Data Output Mode"

The parameter setting of "Data output mode (prt)" applies to the performance when the "Data memory (data)" parameter is set to "2" (to store the weighing data) and when the data is transmitted using the RS-232C interface.

Key mode

When the

 

 

key is pressed with the stabilization indictor turned on, the balance

PRINT

outputs or stores the weighing data and the display blinks one time.

Required setting

dout prt 0

Key mode

Auto print modes A and B

When the displayed value is stable and the conditions of "Auto print polarity", "Auto print difference" and reference value are met, the balance outputs or stores the weighing data. When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or stores the data and the display blinks one time.

Auto print modes A

 

 

 

Example

“For weighing each time a sample is added (or removed), with

 

"ar-d" set to "1" (to adjust zero after the data is output).”

Required setting

dout

prt 1

Auto print mode A (reference = zero)

 

dout

ap-p

Auto print polarity

 

dout

ap-b

Auto print difference

 

dout

ar-d 1

Zero after output

Auto print modes B

 

 

 

Example

For weighing while a sample is added.

Required setting

dout

prt 2

Auto print mode B (reference = last stable value)

 

dout

ap-p

Auto print polarity

 

dout

ap-b

Auto print difference

Stream mode

The balance outputs the weighing data continuously regardless of the display condition. The display does not blink in this mode. The interval memory mode is used when the "Data memory (data)" parameter is set to "2" (to store the weighing data).

Example

For monitoring data on a computer.

Required setting

dout

prt 3

Stream mode / interval memory mode

 

dout

data 0

Data memory function is not used

 

ba5fnc

5pd

Display refresh rate

 

5if

bp5

Baud rate

Caution The balance may not transmit the data completely at the specified refresh rate, depending on the baud rate or data added to the weighing data such as time, date and ID number.

33

Image 35 Contents
1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 14-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 10-510-6 About This Manual FeaturesIntroduction Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Do not allow vibration or drafts to affect Self Check Function with Response AdjustmentBalance during adjustment Terms CalibrationCalibration Group 150 g ~ +0.159 g Automatic self calibration due to changes in temperatureAutomatic Self Calibration About the internal mass Calibration using the internal massOne-Touch Calibration When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output mode Cp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableDisplay character set Setting the ID NumberID Number And GLP Report Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hSignature Space, Ascii 20h Calibration report using an external weightCalibration test report using an external weight End block Title block and end blockOperation Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersOption RS-232C TerminalsExternal contact input Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited