Calibration report using an external weight
Setting of "info 1" AD-8121 printer format
CALIBRATED(EXT.)
CAL.WEIGHT
SIGNATURE
Manufacturer
Model
Serial number
ID number
Date
Time Calibration type
Calibration weight Signature
Setting of "info 2" General format
~~~~~~~~~~~A~&~D<TERM>
~~~~~~2009/12/31<TERM>
TIME<TERM>
~~~~~~~~12:34:56<TERM> CALIBRATED(EXT.)<TERM>
CAL.WEIGHT<TERM>
~~~~+20000.00~~g<TERM>
|
|
|
| Space, ASCII 20h |
|
|
|
| |
<TERM> | Terminator, CR , LF or CR | |||
CR | Carriage return, ASCII 0Dh | |||
LF | Line feed, ASCII 0Ah |
SIGNATURE<TERM> <TERM>
<TERM>
<TERM>
<TERM>
Calibration test report using an external weight
Note Calibration test does not perform calibration.
Setting of "info 1" | Setting of "info 2" |
General format |
Manufacturer | |||
Model | |||
01234567 | Serial number | ||
ABCDEFG | ID number | ||
2009/12/31 | Date | ||
12:34:56 | Time | ||
CAL.TEST(EXT.) |
| Calibration test type | |
ACTUAL |
|
| Zero point value |
g | |||
g | Target weight value | ||
TARGET |
|
| Target weight |
g |
SIGNATURE
Signature
|
|
|
| Space, ASCII 20h |
|
|
|
| |
<TERM> | Terminator, CR , LF or CR | |||
CR | Carriage return, ASCII 0Dh | |||
LF | Line feed, ASCII 0Ah |
~~~~~~~~~~~A~&~D<TERM>
~~~~~~2009/12/31<TERM>
TIME<TERM>
~~~~~~~~12:34:56<TERM> CAL.TEST(EXT.) <TERM>
ACTUAL<TERM>
~~~~~~~~~0.00~~g<TERM>
~~~~+19999.99~~g<TERM>
TARGET<TERM>
~~~~+20000.00~~g<TERM>
SIGNATURE<TERM> <TERM>
<TERM>
<TERM>
<TERM>
40