A&D MC-30K Automatic response adjustment zero error, Automatic response adjustment unstable error

Page 57

Display

Error code

Description

 

 

 

 

 

 

 

Automatic response adjustment zero error

 

 

The automatic response adjustment can not be performed

 

 

because there is something on the pan. Clear the pan. Press

 

 

the

 

key to return to the weighing mode.

 

 

CAL

 

 

Automatic response adjustment unstable error

 

 

The automatic response adjustment can not be performed

 

 

because the weight value is unstable.

(Check NG)

Check the ambient conditions such as breeze, vibration

and magnetic fields, also check the weighing pan.

 

 

 

 

Press the

 

key to return to the weighing mode.

 

 

CAL

 

 

Internal error

(Check no)

This error indicates an internal error as the result of the self-check

function. Repair is required. Contact the local A&D dealer.

 

 

Clock battery error

 

 

The clock backup battery has been depleted.

 

 

Press any key and set the time and date. The clock and

 

 

calendar function works normally as long as the AC adapter

 

 

is connected to the balance. If this error appears frequently,

 

 

contact the local A&D dealer.

 

 

Memory full

 

(Blinking)

The amount of weighing data in memory has reached the

 

maximum capacity. Delete the data in memory to store new

 

 

data. For details, refer to the GX-K manual "12. Data Memory".

 

 

Memory full

 

 

The amount of calibration or calibration test data in memory

 

(Illuminated)

has reached the maximum capacity (50 sets).

 

The data in memory will be deleted automatically to store

 

 

new data. For details, refer to the GX-K manual "12. Data

 

 

Memory".

 

 

Memory type error

 

 

The type of memory set in the function table and the type of

 

 

data stored are different. For details, refer to the GX-K manual

 

 

"12. Data Memory".

 

 

Communications error

 

EC,E00

A protocol error occurred in communications.

 

 

Confirm the format, baud rate and parity.

 

EC,E01

Undefined command error

 

An undefined command was received.

 

 

Confirm the command.

 

 

Not ready

 

 

A received command can not be processed.

 

 

Example:

 

EC,E02

The balance received a "Q" command, but not in the

 

 

weighing mode.

 

 

The balance received a "Q" command while processing a

 

 

RE-ZERO command.

 

 

Adjust the delay time to transmit a command.

 

 

55

Image 57
Contents 1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 10-5 10-614-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Features IntroductionAbout This Manual Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Key When pressed and released When pressed and held Display Symbols And Key OperationKey Operations DisplayWeighing Units Tael HK general, Singapore Weighing Selecting a Weighing Unit ModeBasic Weighing F1l Filter Used DisplayWhen Building into a System Weighing Speed Adjustment Self Check Function with Response Adjustment Balance during adjustmentDo not allow vibration or drafts to affect Calibration Calibration GroupTerms Automatic self calibration due to changes in temperature Automatic Self Calibration150 g ~ +0.159 g Calibration using the internal mass One-Touch CalibrationAbout the internal mass When the MC-10K is used Calibration Using an External WeightPrint key...... To store the new weight value. Even if Calibration Test Using an External WeightRemove Remove the weight from the pan Step Balance displays Step Confirm that there is nothing on the pan and pressCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationInitializing the Balance To cancel This operation, press KeyStep Press Key Step Press Key to displayFunction Table Check the settings and condition before changing parametersStructure and Sequence of the Function Table Display and Operation KeysDetails of the Function Table Cp fnc Cp-b Data output modeBep LO buzzerAK, Error code TimeoutTime/Date output Dout ID number outputInternal mass correction ID number settingF1-b Averaging range for First moving Average F1-tDescription of the Class Environment, Display Zero upon power-on p-trCondition Cond Stability band width 5t-bKey mode Description of the Item Data Output ModeStream mode Auto print modes a and BDescription of the Item Data Format Interval memory modeStandard format 5if type DP Dump print format 5if typeMT format KF formatNU numerical format 5if type CSV formatDate Data numberTime Tare valueData Format Examples Negative errorStable OverloadSetting the ID Number ID Number And GLP ReportDisplay character set GLP Report Calibration report using the internal massSpace, Ascii 20h Terminator, CR , LF or CRCalibration report using an external weight Calibration test report using an external weightSignature Space, Ascii 20h Title block and end block OperationEnd block RS-232C and External Contact Input Standard Input And Output InterfaceSub 25 pin numbers Sub 25 pin assignmentsRS-232C Terminals External contact inputOption Not Auto print polarity SettingsNecessary Data output pauseUsing the WinCT software, the balance can do the following Using Windows Communication Tools Software WinCTRsCom RsKeyCommand List CommandsAcknowledge Code and Error Codes Settings Related to RS-232C Control Using CTS and RTSCommand Personal computer Time Balance Command Weighing modeExtended Function Page Page Treatment of the Balance MaintenanceChecking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Weighing method16-2 Automatic response adjustment unstable error Automatic response adjustment zero errorInternal error Clock battery errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsOptions and Peripheral Instruments AD-8922A Remote controllerAX-MC10K/30KPAN Auto-centering Pan AD-8121B PrinterFoot Switch Rechargeable BatteryWeighing Environment Logger Data LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsMode MT format ErrorF1l FUlPage German Sales Office UK OfficeKorea Limited Instruments India Private Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.