A&D MC-30K manual Calibration Using an External Weight, When the MC-10K is used

Page 23
Place weight
end
Step 1 Step 2

8-4. Calibration Using an External Weight

This function calibrates the balance using an external weight.

When the MC-10K is used

Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan. Press and hold the CAL key until Calout

0 .000 g

Press and hold

Calout

 

 

is displayed, then release the key.

Step

3 The balance displays

Cal 0

.

 

 

If you want to change

the calibration weight,

 

 

press the

 

 

key and go to step 4.

 

SAMPLE

 

 

If you use

the calibration weight value stored

 

 

in the balance, go to step 5.

Step

4 Specify the calibration weight value as follows:

 

 

key... To switch between the calibration weight

SAMPLE

 

 

 

selection mode (All of the digits blinking) or

 

 

 

the value adjustment mode (The selected

 

 

 

digits blinking).

 

 

 

 

 

With the MC-10K, the digits in the 1st, 2nd,

 

 

 

and 3rd decimal places can be adjusted.

 

 

 

With the MC-31K, the digits in the 1st and

 

 

 

2nd decimal places can be adjusted.

 

key.. To select the calibration weight (All of the

RE-ZERO

 

 

 

digits blinking) or adjust the value (The

 

 

 

selected digits blinking). Refer to page 19.

PRINT key...... To store the new weight value. Even if the

 

AC adapter is removed, the data is

 

maintained in non-volatile memory.

 

key ......... To cancel the operation and return to

 

.

CAL

Cal 0

Step

5 Confirm that there is nothing on the pan

and

 

Press

1

0000 .000

Select

2000 .000

2000 .000

2000 .120

2000 .120

2000 .123

Release

Cal 0

g

g

g

g

g

g

Cal 0

 

 

press the

 

key. The balance measures

 

 

PRINT

 

 

the zero point. Do not allow vibration or drafts to

 

 

affect the balance. The balance displays the

 

 

calibration weight value.

Step

6

Place the displayed calibration weight on the

 

 

pan and press the

 

 

 

 

key. The balance

 

 

 

PRINT

 

 

measures the calibration weight. Do not allow

 

 

vibration or drafts to affect the balance.

Step

7

The balance displays

 

 

.

 

end

Remove the weight from the pan.

Example:

2 k

 

New weight

 

2000.123 g

 

Remove

Step

8

If the "GLP output (info)" parameter, of the function table,

 

 

 

 

 

is set to "1" or "2", the balance displays

glp

and

GLP output

glp

 

 

 

outputs the "Calibration Report" using the RS-232C

 

 

 

 

 

 

 

 

 

interface or stores the data in memory. For details on the

 

end

 

 

 

calibration report format, refer to "11-2. GLP Report".

 

 

 

Step

9

The balance will automatically return to the weighing mode.

0

.000 g

Step 10

Place the calibration weight on the pan and confirm that the value

 

 

displayed is within ±20 digits of the specified value. If it is not within the range, check the ambient conditions such as breeze and vibration also check the weighing pan. Then, repeat steps 1 to 10.

21

Image 23
Contents 1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 14-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 10-510-6 About This Manual FeaturesIntroduction Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Do not allow vibration or drafts to affect Self Check Function with Response AdjustmentBalance during adjustment Terms CalibrationCalibration Group 150 g ~ +0.159 g Automatic self calibration due to changes in temperatureAutomatic Self Calibration About the internal mass Calibration using the internal massOne-Touch Calibration When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output modeCp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableDisplay character set Setting the ID NumberID Number And GLP Report Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hSignature Space, Ascii 20h Calibration report using an external weightCalibration test report using an external weight End block Title block and end blockOperation Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersOption RS-232C TerminalsExternal contact input Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.