8-6. Correcting the Internal Mass Value
The balance can correct the internal mass value within the range shown below. This function corrects the internal mass value to conform to an external weight. The corrected mass value is maintained in
internal mass value is corrected as follows:
Model | Target | Range | 10 kg |
2 kg | ±0.500 g | 10000.00 g | |
10 kg | ±5.00 g |
The same mass
Correction the internal mass by +0.30 g at 10 kg. Calibrate with the corrected internal mass.
10 kg
10000.30 g
Corrected external weight
Step 1 Calibrate the balance using the internal mass.
Example for correcting the weight value by +0.30 g in 10 kg, when using
Step 2 Press the
ON:OFF
key to turn off the display.
Step 3 While pressing and holding the PRINT the SAMPLE key, press the ON:OFF balance displays p5 .
key and key. The
With these keys Press the key held down
p5
Step 4 Press the PRINT key. Then the balance displays the function switches. Set the function table switch and internal mass correction switch to "1" as shown above using the following keys. SAMPLE key...To select the switch to change the value.
The selected digit blinks.
Correction switch of internal mass
a - 01111
a - 1 x x x 1
Step | 5 | Press the |
| key to store the new setting. | |||||
|
| The balance returns to the weighing mode. | |||||||
Step | 6 | Press and hold the |
|
| key to enter the function | ||||
| SAMPLE | ||||||||
|
| table and release the key when |
| is displayed. | |||||
|
| ba5fnc |
0 .0 g
Press and hold
ba5fnc
Step 7 Press the SAMPLE key several times until C5 in displayed, then release the key.
Step 8 Press the PRINT key to enter the procedure for correcting the internal mass value.
is
C5 in
Step 9 Correct the internal mass value using the following keys. | 0 .30 g | |
|
| after +500 digits) |
key | ....... |
To store the new value and display the | |
| next menu item of the function table. |
23