A&D MC-30K, MC-10K manual Correcting the Internal Mass Value

Page 25

8-6. Correcting the Internal Mass Value

The balance can correct the internal mass value within the range shown below. This function corrects the internal mass value to conform to an external weight. The corrected mass value is maintained in non-volatile memory even if the AC adapter is removed. The

internal mass value is corrected as follows:

Model

Target

Range

10 kg

MC-10K

2 kg

±0.500 g

10000.00 g

MC-30K

10 kg

±5.00 g

The same mass

Correction the internal mass by +0.30 g at 10 kg. Calibrate with the corrected internal mass.

10 kg

10000.30 g

Corrected external weight

Step 1 Calibrate the balance using the internal mass. (one-touch calibration). Place the external weight and find out the correction value.

Example for correcting the weight value by +0.30 g in 10 kg, when using MC-30K. If correcting the weight value by +0.30 g in 5 kg, the weight changed into 10 kg, the correction value is +0.60 g.

Step 2 Press the

ON:OFF

key to turn off the display.

Step 3 While pressing and holding the PRINT the SAMPLE key, press the ON:OFF balance displays p5 .

key and key. The

With these keys Press the key held down

p5

Step 4 Press the PRINT key. Then the balance displays the function switches. Set the function table switch and internal mass correction switch to "1" as shown above using the following keys. SAMPLE key...To select the switch to change the value.

The selected digit blinks.

RE-ZEROkey ..To change the parameter of the switch selected. Function table switch

Correction switch of internal mass

a - 01111

a - 1 x x x 1

Step

5

Press the

PRINT

 

key to store the new setting.

 

 

The balance returns to the weighing mode.

Step

6

Press and hold the

 

 

key to enter the function

 

SAMPLE

 

 

table and release the key when

 

is displayed.

 

 

ba5fnc

0 .0 g

Press and hold

ba5fnc

Step 7 Press the SAMPLE key several times until C5 in displayed, then release the key.

Step 8 Press the PRINT key to enter the procedure for correcting the internal mass value.

is

C5 in

Step 9 Correct the internal mass value using the following keys.

0 .30 g

RE-ZEROkey..To select the value. (-500 digits appear

 

PRINT

 

after +500 digits)

key

.......

To store the new value and display the

 

next menu item of the function table.

23

Image 25 Contents
1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 10-6 10-514-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Introduction FeaturesAbout This Manual Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Key When pressed and released When pressed and held Display Symbols And Key OperationKey Operations DisplayWeighing Units Tael HK general, Singapore Weighing Selecting a Weighing Unit ModeBasic Weighing F1l Filter Used DisplayWhen Building into a System Weighing Speed Adjustment Balance during adjustment Self Check Function with Response AdjustmentDo not allow vibration or drafts to affect Calibration Group CalibrationTerms Automatic Self Calibration Automatic self calibration due to changes in temperature 150 g ~ +0.159 g One-Touch Calibration Calibration using the internal massAbout the internal mass When the MC-10K is used Calibration Using an External WeightPrint key...... To store the new weight value. Even if Calibration Test Using an External WeightRemove Remove the weight from the pan Step Balance displays Step Confirm that there is nothing on the pan and pressCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationInitializing the Balance To cancel This operation, press KeyStep Press Key Step Press Key to displayFunction Table Check the settings and condition before changing parametersStructure and Sequence of the Function Table Display and Operation KeysDetails of the Function Table Cp fnc Cp-b Data output modeBep LO buzzerAK, Error code TimeoutTime/Date output Dout ID number outputInternal mass correction ID number settingF1-b Averaging range for First moving Average F1-tDescription of the Class Environment, Display Zero upon power-on p-trCondition Cond Stability band width 5t-bKey mode Description of the Item Data Output ModeStream mode Auto print modes a and BDescription of the Item Data Format Interval memory modeStandard format 5if type DP Dump print format 5if typeMT format KF formatNU numerical format 5if type CSV formatDate Data numberTime Tare valueData Format Examples Negative errorStable OverloadID Number And GLP Report Setting the ID NumberDisplay character set GLP Report Calibration report using the internal massSpace, Ascii 20h Terminator, CR , LF or CRCalibration test report using an external weight Calibration report using an external weightSignature Space, Ascii 20h Operation Title block and end blockEnd block RS-232C and External Contact Input Standard Input And Output InterfaceSub 25 pin numbers Sub 25 pin assignmentsExternal contact input RS-232C TerminalsOption Not Auto print polarity SettingsNecessary Data output pauseUsing the WinCT software, the balance can do the following Using Windows Communication Tools Software WinCTRsCom RsKeyCommand List CommandsAcknowledge Code and Error Codes Settings Related to RS-232C Control Using CTS and RTSCommand Personal computer Time Balance Command Weighing modeExtended Function Page Page Treatment of the Balance MaintenanceChecking the Balance Performance and Environment TroubleshootingChecking that the balance performs properly Weighing method16-2 Automatic response adjustment unstable error Automatic response adjustment zero errorInternal error Clock battery errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsOptions and Peripheral Instruments AD-8922A Remote controllerAX-MC10K/30KPAN Auto-centering Pan AD-8121B PrinterFoot Switch Rechargeable BatteryWeighing Environment Logger Data LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsMode MT format ErrorF1l FUlPage German Sales Office UK OfficeKorea Limited Instruments India Private Limited