A&D MC-10K, MC-30K manual Control Using CTS and RTS, Settings Related to RS-232C, Command

Page 50

R command

 

Personal computer Time

Balance

R command

Weighing mode

 

R

CR LF

AKCR LF

AKCR LF

Command received In process

Command completed Zero display

When a communication error has occurred due to external noise, or a parity error has occurred due to transmission error, the balance transmits an error code. In this case, send the command again.

xxis error code number.

13-3. Control Using CTS and RTS

Depending on the "Ct5" parameter of "Serial interface (5if)", the balance performs as follows:

Ct5 0

Regardless of whether the balance can receive a command or not, the balance keeps the CTS line HI. The balance outputs data regardless of the condition of the RTS line.

Ct5 1

The CTS line is kept Hi normally. When the balance can not receive the next command (Example: while the balance is processing the last command), the balance sets the CTS line to Lo. The balance confirms the level of the RTS line before outputting a set of data. If the RTS level is Hi, the balance outputs data. If the RTS level is Lo, data is not output (The data is canceled).

13-4. Settings Related to RS-232C

Concerning the RS-232C, the balance has two functions: "Data output (dout)" and "Serial interface (5if)". Set each function as necessary.

48

Image 50
Contents MC-10K MC-30K 1WMPD4002203AThis Manual and Marks All rights reservedContents 14-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 10-510-6 About This Manual FeaturesIntroduction Compliance Compliance with the following Council DirectivesUnpacking And Installing The Balance To keep waterproof DustproofAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Installing the Balance EarthPrecautions Before UseDuring Use After UsePower Supply Key Operations Display Symbols And Key OperationKey When pressed and released When pressed and held DisplayWeighing Units Tael HK general, Singapore Basic Weighing Selecting a Weighing Unit ModeWeighing F1l Filter Used DisplayWhen Building into a System Weighing Speed Adjustment Do not allow vibration or drafts to affect Self Check Function with Response AdjustmentBalance during adjustment Terms CalibrationCalibration Group 150 g ~ +0.159 g Automatic self calibration due to changes in temperatureAutomatic Self Calibration About the internal mass Calibration using the internal massOne-Touch Calibration Calibration Using an External Weight When the MC-10K is usedRemove Remove the weight from the pan Step Balance displays Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Step Confirm that there is nothing on the pan and pressCorrecting the Internal Mass Value Key Function Switch And Initialization Permit or InhibitStep Press Key To cancel This operation, press KeyInitializing the Balance Step Press Key to displayStructure and Sequence of the Function Table Check the settings and condition before changing parametersFunction Table Display and Operation KeysDetails of the Function Table Bep Data output modeCp fnc Cp-b LO buzzerTime/Date output TimeoutAK, Error code Dout ID number outputF1-b Averaging range for First moving Average ID number settingInternal mass correction F1-tCondition Cond Zero upon power-on p-trDescription of the Class Environment, Display Stability band width 5t-bStream mode Description of the Item Data Output ModeKey mode Auto print modes a and BStandard format 5if type Interval memory modeDescription of the Item Data Format DP Dump print format 5if typeNU numerical format 5if type KF formatMT format CSV formatTime Data numberDate Tare valueStable Negative errorData Format Examples OverloadDisplay character set Setting the ID NumberID Number And GLP Report Space, Ascii 20h Calibration report using the internal massGLP Report Terminator, CR , LF or CRSignature Space, Ascii 20h Calibration report using an external weightCalibration test report using an external weight End block Title block and end blockOperation Sub 25 pin numbers Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin assignmentsOption RS-232C TerminalsExternal contact input Necessary SettingsNot Auto print polarity Data output pauseRsCom Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsKeyCommands Command ListAcknowledge Code and Error Codes Command Control Using CTS and RTSSettings Related to RS-232C Personal computer Time Balance Command Weighing modeExtended Function Page Page Maintenance Treatment of the BalanceChecking that the balance performs properly TroubleshootingChecking the Balance Performance and Environment Weighing method16-2 Internal error Automatic response adjustment zero errorAutomatic response adjustment unstable error Clock battery errorOther Display Asking for RepairSpecifications External Dimensions Unit mmAX-MC10K/30KPAN Auto-centering Pan AD-8922A Remote controllerOptions and Peripheral Instruments AD-8121B PrinterWeighing Environment Logger Rechargeable BatteryFoot Switch Data LoggerTerms/Index TermsMemory function Nothing is on the weighing panKeys and symbols IndexF1l ErrorMode MT format FUlPage Korea Limited UK OfficeGerman Sales Office Instruments India Private Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.