A&D MC-30K Troubleshooting, Checking the Balance Performance and Environment, Weighing method

Page 55

16. Troubleshooting

16-1. Checking the Balance Performance and Environment

The balance is a precision instrument. When the operating environment or the operating method is inadequate, correct weighing can not be performed. Place a sample on the pan and remove it, and repeat this several times. if the balance seems to have a problem with repeatability or to perform improperly, check as described below. If improper performance persists after checking, contact the local A&D dealer for repair.

Checking that the balance performs properly

Check the balance performance using the self-check function as described in "7. Weighing Speed Adjustment / Self Check Function".

An error display appears when a malfunction is found.

Check the balance repeatability using an external weight. Be sure to place the weight in the center of the weighing pan.

Check the balance repeatability, linearity and calibrated value using external weights with a known value.

Checking that the operating environment or weighing method is proper Operating environment

Is the weighing table solid enough?

Is the balance level? Refer to "3-1. Before Use".

Is the operating environment free from vibration and drafts?

Is there a strong electrical or magnetic noise source such as a motor near the balance?

Weighing method

Does the weighing pan touch the rim or anything? Is the weighing pan and pan support installed correctly?

Is the RE-ZEROkey pressed before placing a sample on the weighing pan? Is the sample placed in the center of the weighing pan?

Has the balance been calibrated using the internal mass (one-touch calibration)? Has the balance been warmed up for 30 minutes before weighing?

Sample and container

Has the sample absorbed or lost moisture due to the ambient conditions such as temperature and humidity?

Has the temperature of the container been allowed to equalize to the ambient temperature? Refer to "3-2. During Use".

Is the sample charged with static electricity? Refer to "3-2. During Use".

Is the sample of magnetic material such as iron? Caution is required for weighing magnetic materials. Refer to "3-2. During Use".

53

Image 55
Contents 1WMPD4002203A MC-10K MC-30KAll rights reserved This Manual and MarksContents 10-6 10-514-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Introduction FeaturesAbout This Manual Compliance with the following Council Directives ComplianceTo keep waterproof Dustproof Unpacking And Installing The BalanceAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Earth Installing the BalanceBefore Use PrecautionsAfter Use During UsePower Supply Display Display Symbols And Key OperationKey When pressed and released When pressed and held Key OperationsWeighing Units Tael HK general, Singapore F1l Filter Used Display Selecting a Weighing Unit ModeWeighing Basic WeighingWhen Building into a System Weighing Speed Adjustment Balance during adjustment Self Check Function with Response AdjustmentDo not allow vibration or drafts to affect Calibration Group CalibrationTerms Automatic Self Calibration Automatic self calibration due to changes in temperature150 g ~ +0.159 g One-Touch Calibration Calibration using the internal massAbout the internal mass When the MC-10K is used Calibration Using an External WeightStep Confirm that there is nothing on the pan and press Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Remove Remove the weight from the pan Step Balance displaysCorrecting the Internal Mass Value Key Permit or Inhibit Function Switch And InitializationStep Press Key to display To cancel This operation, press KeyInitializing the Balance Step Press KeyDisplay and Operation Keys Check the settings and condition before changing parametersFunction Table Structure and Sequence of the Function TableDetails of the Function Table LO buzzer Data output modeCp fnc Cp-b BepDout ID number output TimeoutAK, Error code Time/Date outputF1-t ID number settingInternal mass correction F1-b Averaging range for First moving AverageStability band width 5t-b Zero upon power-on p-trDescription of the Class Environment, Display Condition CondAuto print modes a and B Description of the Item Data Output ModeKey mode Stream modeDP Dump print format 5if type Interval memory modeDescription of the Item Data Format Standard format 5if typeCSV format KF formatMT format NU numerical format 5if typeTare value Data numberDate TimeOverload Negative errorData Format Examples StableID Number And GLP Report Setting the ID NumberDisplay character set Terminator, CR , LF or CR Calibration report using the internal massGLP Report Space, Ascii 20hCalibration test report using an external weight Calibration report using an external weightSignature Space, Ascii 20h Operation Title block and end blockEnd block Sub 25 pin assignments Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin numbersExternal contact input RS-232C TerminalsOption Data output pause SettingsNot Auto print polarity NecessaryRsKey Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsComCommand List CommandsAcknowledge Code and Error Codes Personal computer Time Balance Command Weighing mode Control Using CTS and RTSSettings Related to RS-232C CommandExtended Function Page Page Treatment of the Balance MaintenanceWeighing method TroubleshootingChecking the Balance Performance and Environment Checking that the balance performs properly16-2 Clock battery error Automatic response adjustment zero errorAutomatic response adjustment unstable error Internal errorAsking for Repair Other DisplaySpecifications Unit mm External DimensionsAD-8121B Printer AD-8922A Remote controllerOptions and Peripheral Instruments AX-MC10K/30KPAN Auto-centering PanData Logger Rechargeable BatteryFoot Switch Weighing Environment LoggerTerms Terms/IndexNothing is on the weighing pan Memory functionIndex Keys and symbolsFUl ErrorMode MT format F1lPage Instruments India Private Limited UK OfficeGerman Sales Office Korea Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.