A&D MC-10K One-Touch Calibration, Calibration using the internal mass, About the internal mass

Page 22

8-3. One-Touch Calibration

Calibration using the internal mass

This function calibrates the balance using the internal mass. The only operation required is

to press the

CAL

key.

Step 1 Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan.

Step 2 Press the

CAL

key.

Step

3

The balance displays

Calin

and performs calibration using the internal mass. Do

 

 

not allow vibration or

drafts to

affect the balance.

Step

4

The balance displays

 

 

after calibration. If the "GLP output (info)" parameter of

 

end

 

 

the function table is set to "1" or "2", the balance displays

 

and outputs the

 

 

glp

 

 

"calibration report" using the RS-232C interface or stores

the data in memory. Refer

 

 

to "11-2. GLP Report" and "Data memory (data)" of the function table for details.

Step

5

The balance will automatically return to the weighing mode after calibration.

About the internal mass

The value of the internal mass may change due to aging, corrosion or other damage caused by the operating environment. Check the internal mass periodically. Correct the internal mass value as necessary. Refer to "8-6. Correcting the Internal Mass Value".

To maintain the weighing accuracy, perform the calibration using an external weight periodically, as described below.

20

Image 22
Contents MC-10K MC-30K 1WMPD4002203AThis Manual and Marks All rights reservedContents 10-6 10-514-1-1 Averaging Range f1-b and Averaging Time f1-t 14-1-2 Introduction FeaturesAbout This Manual Compliance Compliance with the following Council DirectivesUnpacking And Installing The Balance To keep waterproof DustproofAuto-centering Pan AX-MC10K/30KPAN Installation Procedure Installing the Balance EarthPrecautions Before UseDuring Use After UsePower Supply Key Operations Display Symbols And Key OperationKey When pressed and released When pressed and held DisplayWeighing Units Tael HK general, Singapore Basic Weighing Selecting a Weighing Unit ModeWeighing F1l Filter Used DisplayWhen Building into a System Weighing Speed Adjustment Balance during adjustment Self Check Function with Response AdjustmentDo not allow vibration or drafts to affect Calibration Group CalibrationTerms Automatic Self Calibration Automatic self calibration due to changes in temperature150 g ~ +0.159 g One-Touch Calibration Calibration using the internal massAbout the internal mass Calibration Using an External Weight When the MC-10K is usedRemove Remove the weight from the pan Step Balance displays Calibration Test Using an External WeightPrint key...... To store the new weight value. Even if Step Confirm that there is nothing on the pan and pressCorrecting the Internal Mass Value Key Function Switch And Initialization Permit or InhibitStep Press Key To cancel This operation, press KeyInitializing the Balance Step Press Key to displayStructure and Sequence of the Function Table Check the settings and condition before changing parametersFunction Table Display and Operation KeysDetails of the Function Table Bep Data output modeCp fnc Cp-b LO buzzerTime/Date output TimeoutAK, Error code Dout ID number outputF1-b Averaging range for First moving Average ID number settingInternal mass correction F1-tCondition Cond Zero upon power-on p-trDescription of the Class Environment, Display Stability band width 5t-bStream mode Description of the Item Data Output ModeKey mode Auto print modes a and BStandard format 5if type Interval memory modeDescription of the Item Data Format DP Dump print format 5if typeNU numerical format 5if type KF formatMT format CSV formatTime Data numberDate Tare valueStable Negative errorData Format Examples OverloadID Number And GLP Report Setting the ID NumberDisplay character set Space, Ascii 20h Calibration report using the internal massGLP Report Terminator, CR , LF or CRCalibration test report using an external weight Calibration report using an external weightSignature Space, Ascii 20h Operation Title block and end blockEnd block Sub 25 pin numbers Standard Input And Output InterfaceRS-232C and External Contact Input Sub 25 pin assignmentsExternal contact input RS-232C TerminalsOption Necessary SettingsNot Auto print polarity Data output pauseRsCom Using Windows Communication Tools Software WinCTUsing the WinCT software, the balance can do the following RsKeyCommands Command ListAcknowledge Code and Error Codes Command Control Using CTS and RTSSettings Related to RS-232C Personal computer Time Balance Command Weighing modeExtended Function Page Page Maintenance Treatment of the BalanceChecking that the balance performs properly TroubleshootingChecking the Balance Performance and Environment Weighing method16-2 Internal error Automatic response adjustment zero errorAutomatic response adjustment unstable error Clock battery errorOther Display Asking for RepairSpecifications External Dimensions Unit mmAX-MC10K/30KPAN Auto-centering Pan AD-8922A Remote controllerOptions and Peripheral Instruments AD-8121B PrinterWeighing Environment Logger Rechargeable BatteryFoot Switch Data LoggerTerms/Index TermsMemory function Nothing is on the weighing panKeys and symbols IndexF1l ErrorMode MT format FUlPage Korea Limited UK OfficeGerman Sales Office Instruments India Private Limited

MC-10K, MC-30K specifications

The A&D MC-30K and MC-10K are highly regarded weighing scales that cater to diverse industrial and commercial applications. Renowned for their accuracy, durability, and advanced technological features, these models are ideal for users who require precision in measurement.

The MC-30K model offers a capacity of up to 30 kilograms, while the MC-10K is designed for loads of up to 10 kilograms. Both models are equipped with high-resolution load cells that provide outstanding accuracy and stability in measurements. The readability of 1 gram for the MC-10K and 5 grams for the MC-30K ensures that even the smallest variations in weight are captured effectively, making them suitable for activities ranging from laboratory analysis to quality control in manufacturing processes.

One of the standout features of these scales is their easy-to-read LCD display, which allows users to view measurements clearly, even in low-light conditions. The display is designed to provide quick and efficient readouts, minimizing wait times during the weighing process.

Both the MC-30K and MC-10K models incorporate advanced technology, including automatic calibration and tare functions, which enhance usability and accuracy. The automatic calibration feature ensures that the scales maintain their precision over time, while the tare function allows users to weigh items in containers without needing to account for the container's weight separately.

Durability is another major characteristic of these scales. Built with robust materials, they are designed to withstand the rigors of daily use in busy environments. Additionally, their compact design ensures that they occupy minimal space, making them suitable for various settings, including laboratories, warehouses, and retail outlets.

Convenience is amplified with the inclusion of RS-232 interface options for seamless connectivity to computers or printers, enabling easy data collection and reporting. The power-saving features of both models, including auto-off functionality, contribute to longer battery life, reducing the need for frequent replacements.

In summary, the A&D MC-30K and MC-10K scales are exceptional tools for anyone needing reliable and precise weighing solutions. With their advanced features, user-friendly design, and durable construction, they stand out as top choices for industrial and commercial weighing applications.