10-4. Description of the Class "Environment, Display"
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a weight value. Used for powder target weighing, weighing a very light sample or when quick response weighing is required. After setting, the balance displays FAST.
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a weight value from drifting due to vibration or drafts. After setting, the balance displays SLOW.
Notes In automatic response adjustment, the weighing speed is selected automatically.
With "Hold function (Hold)" set to "ON (1)", this item is used to set the averaging time.
Stability band width ( 5t-b )
This item controls the width to regard a weight value as a stable value. When the fluctuation per second is less than the parameter, the balance displays the stabilization indicator and outputs or stores the data. The parameter influences the "Auto print mode"
for exact weighing.
Note | With "Hold function (Hold)" set to "ON (1)", this item is used to set the |
| stabilization range. |
Zero upon power-on ( p-tr )
When a hopper is attached to the weighing pan and
When "
Span calibration ( 5pn )
When a hopper is attached to the weighing pan and calibration is to be performed with the hopper attached, set " 5pn " to “ 1 ”. When the tare value (hopper and other devices attached) is within the value in the table below, calibration using the internal mass is possible.
Model | Tare value |
9 kg or less | |
29 kg or less |