Emerson 1056 Programming the ANALYZER-TUTORIAL, Calibrate, Outputs, Range, O1 S1 4mA 0.000 ppm

Page 25

MODEL FCL-1056

SECTION 4.0

 

DISPLAY AND OPERATION

4.3 PROGRAMMING THE ANALYZER—TUTORIAL

Setting up and calibrating the analyzer is easy. The following tutorial describes how to move around in the programming menus. For practice, the tutorial also describes how to assign ppm chlorine values to the 4 and 20 mA

analog outputs.

Menu

Calibrate

Hold

Program

Display

1.Press MENU. The main menu screen appears. There are four items in the main menu. Calibrate is in reverse video, meaning that the cursor is on Calibrate.

2.To assign values to the analog outputs, the Program sub-menu must be open. Use the down navigation key to move the cursor to Program. Press ENTER.

Program

Outputs

Alarms

Measurement

Temperature

Outputs

Range

Configure

Stimulate

Output Range

O1 S1 4mA 0.000 ppm

O1 S1 20mA: 10.00 ppm

O2

S1 4mA:

0.0C

O2

S1 20mA:

100.0C

O1 S1 20 mA

1 0.00 ppm

3.The Program menu appears. There are between five and seven items in the Program menu. Diagnostic Setup appears only if you have the FCL- 02 with pH sensor. The screen displays four items at a time. The down- ward pointing arrow on the right of the screen shows there are more items available in the menu. To view the other items, use the down key to scroll to the last item shown and continue scrolling down. When you have reached the bottom, the arrow will point up. Move the cursor back to Outputs and press ENTER.

4.The screen at left appears. The cursor is on Range. Output Range is used to assign values to the low and high current outputs. Press ENTER.

5.The screen at left appears. The screen shows the present values assigned to output 1 (O1) and output 2 (O2). The screen also shows which sensors the outputs are assigned to. S1 is sensor 1 and S2 is sensor 2. The assignments shown are the defaults for the FCL-01. For the FCL-02, Output 2 is assigned to sensor 2 (pH). Outputs are freely assignable under the configure menu.

6.For practice, change the 20 mA setting for output 1 to 8.5 ppm.

a.Move the cursor to the O1 S1 20 mA: 10.00 line and press ENTER.

b.The screen at left appears.

c.Use the navigation keys to change 10.00 to 8.5 ppm. Use the left and right keys to move from digit to digit. Use the up and down keys to increase or decrease the numeral.

d.To move the decimal point, press the left or right navigation key until the decimal point is highlighted. Press the up key to move the decimal point to the right. Press the down key to move to the left.

e.Press ENTER to store the setting.

15

Image 25
Contents FCL with 1056 Analyzer Essential Instructions Free Chlorine EnglishPpm mg/L ORPLive/Continous S1 Free Cl PH Correction ManualS1 Manual pH Temp Units Calibrate Sensor 1 Free chlorine Menu TreeSensor 2 pH Temperature AlarmsDiagnostic Setup Reset AnalyzerRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DFeatures Section Description and SpecificationsApplications Specifications Sensor Specifications GeneralSpecifications Analyzer Ordering Information AccessoriesComponent Parts PH Correction required selectionThis page left blank intentionally Unpacking and Inspection Section InstallationFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstalling the Sensors InstallationGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally POWER, ALARM, and Output Wiring Section WiringPower Analog output wiringAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Calibrate Programming the ANALYZER-TUTORIALOutputs RangeO1 S1 20mA 08.50 ppm Output Range O1 S1 4mA 0.000 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityHold Using HoldMain Format Configuring the Main DisplayLanguage English Contrast Format, Language, Warning, and ContrastSection Programming the Analyzer Default SettingsGeneral Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsProgramOutput ConfigureAssign Procedure Ranging Outputs Configure Simulate Output ConfigureOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfiguring Alarms and Assigning Setpoints Configure SimulateRangeSimulate OutputSection Programming the Analyzer Configure/Setpoint AlarmsAlarm Alarm Simulate Alarm Alarms Configure/SetpointSimulate Don’t SimulateSynch Timers Yes Alarms Configure/Setpoint SimulateProcedure Synchronizing Timers Definitions Chlorine Configuring the MeasurementDefinitions pH/ORP Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsTemperature Program Outputs Alarms MeasurementConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompSetting UP Diagnostics Program Alarms Measurement TemperatureSecurity SecurityDiagnostic Setup Program Measurement Temperature SecurityGI Fault High 1500MΩ Procedure Setting Up DiagnosticsProgram Temperature Security Diagnostics Resetting the AnalyzerReset Analyzer Reset AnalyzerSection Calibration Calibrating TemperatureIntroduction Calibrate SensorSensor11 Output ProcedureS1 Calibration S1 Calibration + 25.0CCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalSensor Sensor OutputCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCal10.01 pH 07.01 pHStable Time Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer StandardizeStandardize Redox Temperature S2 pH Cal Buffer CalSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset Output Calibration Analog OutputsMA Output Cal Meter 000 mA Trim Complete This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceReplacing the electrolyte solution and membrane Cleaning the membraneChlorine Sensor GeneralOther Maintenance Cleaning the SensorPH Sensor Constant Head Flow Controller Cleaning the flow controllerWeight Weight Using the Diagnostic Feature Section TroubleshootingFaults OverviewFault message Explanation Section Troubleshooting When a Fault Message is ShowingHardware Error Sensor ADC Error Sensor CPU ErrorSensor Incompatible Sensor RTD OpenGlass Z Too High Sensor RTD Out of RangeReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Process readings are erratic Sensor can be calibrated, but the current is too lowReadings drift Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Simulating Inputs Chlorine Troubleshooting When no Error Message is Showing GeneralAnod CathSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage ASIA-PACIFIC Americas HeadquartersEurope GermanyWarranty Specifications subject to change without notice