Emerson PN 51-FCL-1056 Sensor RTD Out of Range, Glass Z Too High, Reference Impedance Too High

Page 72

MODEL FCL-1056

SECTION 9.0

 

TROUBLESHOOTING

9.3.10 Sensor RTD Out of Range

Both the chlorine and pH sensor contain a Pt 100 RTD (resistance temperature device) for measuring temperature. If the measured resistance is outside the expected range, the analyzer will display the out of range error message.

1.Check wiring connections.

2.Disconnect the sensor from the cable and use an ohmmeter to check the resistance across the RTD. See Figures 9-1 and 9-2. The resistance should be about 110 Ω. If there is an open or short circuit, the sensor has failed and should be replaced. If the resistance is acceptable, attach the sensor to the Variopol cable and disconnect the RTD IN and RTD RTN leads at the analyzer. Connect an ohmmeter across the leads and measure the resistance. If the circuit is open or shorted, the cable must be replaced.

3.If there is no open or short, check the analyzer. See Section 9.10.2.

9.3.11 Glass Z Too High

The sensing element in the pH sensor is a thin glass membrane. Normally, the impedance of the glass membrane is about 80-100 MΩ. As the glass membrane ages, the impedance increases. A large increase in glass impedance suggests that the sensor is near the end of its useful life.

9.3.12 Reference Impedance Too High

The Model 3900VP pH sensor supplied with the FCL-02 has a porous reference junction, so the normal reference impedance is low, less than 5 kΩ. High reference impedance suggests that the junction is severely fouled, the fill solution has become depleted, or the junction is not fully submerged in the sample.

1.Confirm that sample is flowing to the pH flow cell.

2.Clean the reference junction.

3.Check the sensor in buffers. If readings are accurate and the response is reasonably rapid (<5 minutes to reach a stable reading), the sensor is usable. Clear the fault by increasing the reference impedance fault limit. See Section 5.8.2.

4.Replace the sensor if the response in buffers is bad.

9.3.13 Broken Glass

The sensing element in the pH sensor is a thin glass membrane. Normally, the impedance of the glass membrane is about 80-100 MΩ. If the glass membrane gets broken or cracked, the impedance will drop to less than 10 MΩ.

1.Check sensor settings under the Measurement submenu. Confirm that the pre-amplifier location is set to “analyzer”.

2.Confirm that the pH sensor is installed in the flow cell and sample is flowing through the cell.

3.Check the sensor response in two buffers having different pH values. If the membrane is cracked or broken, the pH reading will be about the same in both buffers.

4.Replace the pH sensor.

62

Image 72
Contents FCL with 1056 Analyzer Essential Instructions English Free ChlorinePpm mg/L ORPS1 Free Cl PH Correction Manual Live/ContinousS1 Manual pH Temp Units Menu Tree Calibrate Sensor 1 Free chlorineSensor 2 pH Alarms TemperatureDiagnostic Setup Reset AnalyzerAbout This Document Rev. Level DateTable of Contents FCL-1056Table of Contents CONT’D List of TablesList of Tables CONT’D IiiSection Description and Specifications FeaturesApplications Specifications General Specifications SensorSpecifications Analyzer Accessories Ordering InformationComponent Parts PH Correction required selectionThis page left blank intentionally Section Installation Unpacking and InspectionFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstallation Installing the SensorsGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally Section Wiring POWER, ALARM, and Output WiringPower Analog output wiringSensor Wiring Alarm relay connectionsWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Section Display and Operation DisplayKeypad Programming the ANALYZER-TUTORIAL CalibrateOutputs RangeOutput Range O1 S1 4mA 0.000 ppm O1 S1 20mA 08.50 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityUsing Hold HoldConfiguring the Main Display Main FormatLanguage English Contrast Format, Language, Warning, and ContrastDefault Settings Section Programming the AnalyzerGeneral Alarms Sensor assignmentChoices Default CONFIGURING, RANGING, and Simulating Outputs DefinitionsConfigure ProgramOutputAssign Configure Simulate Output Configure Procedure Ranging OutputsOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfigure Simulate Configuring Alarms and Assigning SetpointsRangeSimulate OutputSection Programming the Analyzer Alarms Configure/SetpointAlarm Alarms Configure/Setpoint Alarm Simulate AlarmSimulate Don’t SimulateAlarms Configure/Setpoint Simulate Synch Timers YesProcedure Synchronizing Timers Configuring the Measurement Definitions ChlorineDefinitions pH/ORP Measurement Sensor1SensorConfiguring Temperature Related Settings Definitions pHProgram Outputs Alarms Measurement TemperatureConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompProgram Alarms Measurement Temperature Setting UP DiagnosticsSecurity SecurityProgram Measurement Temperature Security Diagnostic SetupGI Fault High 1500MΩ Procedure Setting Up DiagnosticsResetting the Analyzer Program Temperature Security DiagnosticsReset Analyzer Reset AnalyzerCalibrating Temperature Section CalibrationIntroduction Procedure Calibrate SensorSensor11 OutputS1 Calibration S1 Calibration + 25.0CProcedure-Zeroing the Sensor Calibration Free ChlorineCalibrate SensorSensor11 Output S1 Calibration Choose Free ChlorineTemperature S1 Calibration ZeroZeroCalCal Process Cal Calibrate?Sensor Sensor OutputCalibration pH Calibration Slope and OffsetZeroBufferCalCal Auto07.01 pH 10.01 pHStable Time Manual BufferManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer Redox Temperature S2 pH Cal Buffer Cal StandardizeStandardizeSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset Calibration Analog Outputs OutputMA Output Cal Meter 000 mA Trim Complete This page left blank intentionally Section Digital Communications Model FCL-1056 Section Section Maintenance AnalyzerCleaning the membrane Replacing the electrolyte solution and membraneChlorine Sensor GeneralCleaning the Sensor Other MaintenancePH Sensor Cleaning the flow controller Constant Head Flow ControllerWeight Weight Section Troubleshooting Using the Diagnostic FeatureFaults OverviewTroubleshooting When a Fault Message is Showing Fault message Explanation SectionHardware Error Sensor CPU Error Sensor ADC ErrorSensor Incompatible Sensor RTD OpenSensor RTD Out of Range Glass Z Too HighReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Sensor can be calibrated, but the current is too low Process readings are erraticReadings drift Sensor does not respond to changes in chlorine level Chlorine readings are too lowTroubleshooting When no Error Message is Showing pH Calibration Error During Two-Point CalibrationCalibration Error during Standardization Sensor Does Not Respond to Known pH ChangesPH Readings Are Moderately Noisy and Tend to Wander Troubleshooting When no Error Message is Showing General Simulating Inputs ChlorineAnod CathSimulating Inputs pH Simulating pH inputSimulating Inputs Temperature Simulating temperaturePage Americas Headquarters ASIA-PACIFICEurope GermanyWarranty Specifications subject to change without notice