Emerson PN 51-FCL-1056 Resetting the Analyzer, Program Temperature Security Diagnostics

Page 46

MODEL FCL-1056

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.9RESETTING THE ANALYZER

5.9.1 Purpose

This section describes how to clear user-entered values and restore default settings. There are three resets:

1.Resetting to factory default values clears ALL user entered settings, including sensor and analog output calibration, and returns ALL settings and calibration values to the factory defaults.

2.Resetting a sensor calibration to the default values clears user-entered calibration data for the selected sensor but leaves all other user-entered data unaffected.

3.Resetting the analog output calibration clears only the user-entered analog output calibration. It leaves all other user-entered settings unchanged.

5.9.2Procedure – Resetting the Analyzer

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

Program

Temperature

Security

Diagnostics

Reset Analyzer

2.Scroll to the bottom of the screen and continue scrolling until Reset Analyzer is highlighted. Press ENTER.

Reset Analyzer

FactoryFactoryDefaultsDefaults

Sensor Cal Only

Output Cal Only

3.Choose whether to reset all user-entered values (Factory Defaults), sensor calibration (Sensor Cal Only), or output calibration (Output Cal Only). If you choose Sensor Cal Only or Output Cal Only a second screen appears in which you can select which sensor or output calibration to reset.

4. To return to the main display, press MENU then EXIT.

36

Image 46
Contents FCL with 1056 Analyzer Essential Instructions Ppm mg/L EnglishFree Chlorine ORPLive/Continous S1 Free Cl PH Correction ManualS1 Manual pH Temp Units Calibrate Sensor 1 Free chlorine Menu TreeSensor 2 pH Diagnostic Setup AlarmsTemperature Reset AnalyzerAbout This Document Rev. Level DateTable of Contents FCL-1056Table of Contents CONT’D List of TablesList of Tables CONT’D IiiFeatures Section Description and SpecificationsApplications Specifications Sensor Specifications GeneralSpecifications Analyzer Component Parts AccessoriesOrdering Information PH Correction required selectionThis page left blank intentionally FCL-01 free chlorine without continuous pH correction Section InstallationUnpacking and Inspection FCL-02 free chlorine with continuous pH correctionGeneral Information InstallationInstalling the Sensors Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally Power Section WiringPOWER, ALARM, and Output Wiring Analog output wiringSensor Wiring Alarm relay connectionsWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Section Display and Operation DisplayKeypad Outputs Programming the ANALYZER-TUTORIALCalibrate RangeO2 S1 4mA 0.0C O2 S1 20mA 100.0C Output Range O1 S1 4mA 0.000 ppmO1 S1 20mA 08.50 ppm SecurityUsing Hold HoldLanguage English Contrast Configuring the Main DisplayMain Format Format, Language, Warning, and ContrastSection Programming the Analyzer Default SettingsGeneral Alarms Sensor assignmentChoices Default CONFIGURING, RANGING, and Simulating Outputs DefinitionsProgramOutput ConfigureAssign Output Output Range Configure Simulate Output ConfigureProcedure Ranging Outputs O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CRangeSimulate Configure SimulateConfiguring Alarms and Assigning Setpoints OutputSection Programming the Analyzer Configure/Setpoint AlarmsAlarm Simulate Alarms Configure/SetpointAlarm Simulate Alarm Don’t SimulateSynch Timers Yes Alarms Configure/Setpoint SimulateProcedure Synchronizing Timers Definitions Chlorine Configuring the MeasurementDefinitions pH/ORP Measurement Sensor1SensorConfiguring Temperature Related Settings Definitions pH Configuring Security Settings Program Outputs Alarms Measurement Temperature S1 Temp Comp Auto S2 Temp CompSecurity Program Alarms Measurement TemperatureSetting UP Diagnostics SecurityGI Fault High 1500MΩ Program Measurement Temperature SecurityDiagnostic Setup Procedure Setting Up DiagnosticsReset Analyzer Resetting the AnalyzerProgram Temperature Security Diagnostics Reset AnalyzerSection Calibration Calibrating TemperatureIntroduction S1 Calibration ProcedureCalibrate SensorSensor11 Output S1 Calibration + 25.0CProcedure-Zeroing the Sensor Calibration Free ChlorineCalibrate SensorSensor11 Output S1 Calibration Choose Free ChlorineSensor Temperature S1 Calibration ZeroZeroCalCal Process CalCalibrate? Sensor OutputCalibration pH Calibration Slope and OffsetZeroBufferCalCal Auto10.01 pH 07.01 pHStable Time Manual BufferManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer Slope 56.19 mV/pH Offset Redox Temperature S2 pH Cal Buffer CalStandardizeStandardize S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset Output Calibration Analog OutputsMA Output Cal Meter 000 mA Trim Complete This page left blank intentionally Section Digital Communications Model FCL-1056 Section Section Maintenance AnalyzerChlorine Sensor Cleaning the membraneReplacing the electrolyte solution and membrane GeneralOther Maintenance Cleaning the SensorPH Sensor Cleaning the flow controller Constant Head Flow ControllerWeight Weight Faults Section TroubleshootingUsing the Diagnostic Feature OverviewFault message Explanation Section Troubleshooting When a Fault Message is ShowingHardware Error Sensor Incompatible Sensor CPU ErrorSensor ADC Error Sensor RTD OpenReference Impedance Too High Sensor RTD Out of RangeGlass Z Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Process readings are erratic Sensor can be calibrated, but the current is too lowReadings drift Sensor does not respond to changes in chlorine level Chlorine readings are too lowTroubleshooting When no Error Message is Showing pH Calibration Error During Two-Point CalibrationCalibration Error during Standardization Sensor Does Not Respond to Known pH ChangesPH Readings Are Moderately Noisy and Tend to Wander Anod Troubleshooting When no Error Message is Showing GeneralSimulating Inputs Chlorine CathSimulating Inputs pH Simulating pH inputSimulating Inputs Temperature Simulating temperaturePage Europe Americas HeadquartersASIA-PACIFIC GermanyWarranty Specifications subject to change without notice