Emerson PN 51-FCL-1056 instruction manual Measurement, Sensor1Sensor

Page 41

MODEL FCL-1056

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.RESOLUTION. The pH display resolution is user selectable: XX.X or XX.XX.

6.FILTER. The analyzer applies a software filter to the raw voltage value coming from the pH sensor. The filter reduces noise, but increases the response time. See section 5.5.2 for more information.

7.REFERENCE IMPEDANCE. Usually the impedance of the reference electrode in a pH sensor is low. However, a few pH sensors have high reference impedance, and the analyzer must be told that the reference impedance is high. The pH sensor used in the FCL-02 has low reference impedance.

5.5.4Procedure – Configuring the Measurement

1.Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER.

Program

Outputs

Alarms

Measurement

Temperature

2.Choose Measurement.

Configure?

Sensor1Sensor 1

Sensor 2

S1 Configure

MeasureMeasure: Free Chlorine

Units: ppm

Filter: 5 sec

Free Cl Correct Live/C

3.The screen at left appears only if you have an FCL-02. Choose Sensor 1 (chlorine) or Sensor 2 (pH).

4.The screen summarizes the present configuration for sensor 1 (chlorine). If you have an FCL-02 the items are Measure, Units, Filter, Free Cl Correct, and Resolution. If you have an FCL-01, the items are Measure, Units, Filter, Manual pH, and Resolution. Only four items are shown at a time. To view the remaining items, scroll to the bottom of the screen and continue scrolling.

To make a change, move the cursor to the desired line and press ENTER. A screen will appear in which the present setting can be edited. To store the setting press ENTER.

a.For Measurement choose Free Chlorine. Do not choose pH

Independ. Free Cl.

b.Leave Filter at the default value (5 sec) unless readings are noisy.

c.If you have an FCL-02, choose either Live/Continuous or Manual for Free Cl Correct (free chlorine correction). Live/Continuous means the analyzer will use the pH measured on the second channel to continuously correct the chlorine reading for changes in the sample pH. Manual means the analyzer will use a fixed pH value entered by the user to convert the raw chlorine signal to a ppm reading.

d.If you have an FCL-01, Free Cl Correct (free chlorine correction) will not appear. Instead, enter the desired pH correction value for Manual pH.

31

Image 41
Contents FCL with 1056 Analyzer Essential Instructions Free Chlorine EnglishPpm mg/L ORPS1 Manual pH Temp Units S1 Free Cl PH Correction ManualLive/Continous Sensor 2 pH Menu TreeCalibrate Sensor 1 Free chlorine Temperature AlarmsDiagnostic Setup Reset AnalyzerRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DApplications Section Description and SpecificationsFeatures Specifications Analyzer Specifications GeneralSpecifications Sensor Ordering Information AccessoriesComponent Parts PH Correction required selectionThis page left blank intentionally Unpacking and Inspection Section InstallationFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstalling the Sensors InstallationGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally POWER, ALARM, and Output Wiring Section WiringPower Analog output wiringAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Calibrate Programming the ANALYZER-TUTORIALOutputs RangeO1 S1 20mA 08.50 ppm Output Range O1 S1 4mA 0.000 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityHold Using HoldMain Format Configuring the Main DisplayLanguage English Contrast Format, Language, Warning, and ContrastGeneral Default SettingsSection Programming the Analyzer Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsAssign ConfigureProgramOutput Procedure Ranging Outputs Configure Simulate Output ConfigureOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfiguring Alarms and Assigning Setpoints Configure SimulateRangeSimulate OutputSection Programming the Analyzer Alarm AlarmsConfigure/Setpoint Alarm Simulate Alarm Alarms Configure/SetpointSimulate Don’t SimulateProcedure Synchronizing Timers Alarms Configure/Setpoint SimulateSynch Timers Yes Definitions pH/ORP Configuring the MeasurementDefinitions Chlorine Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsTemperature Program Outputs Alarms MeasurementConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompSetting UP Diagnostics Program Alarms Measurement TemperatureSecurity SecurityDiagnostic Setup Program Measurement Temperature SecurityGI Fault High 1500MΩ Procedure Setting Up DiagnosticsProgram Temperature Security Diagnostics Resetting the AnalyzerReset Analyzer Reset AnalyzerIntroduction Calibrating TemperatureSection Calibration Calibrate SensorSensor11 Output ProcedureS1 Calibration S1 Calibration + 25.0CCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalSensor Sensor OutputCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCalStable Time 07.01 pH10.01 pH Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer StandardizeStandardize Redox Temperature S2 pH Cal Buffer CalSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset MA Output Cal Meter 000 mA Trim Complete Calibration Analog OutputsOutput This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceReplacing the electrolyte solution and membrane Cleaning the membraneChlorine Sensor GeneralPH Sensor Cleaning the SensorOther Maintenance Constant Head Flow Controller Cleaning the flow controllerWeight Weight Using the Diagnostic Feature Section TroubleshootingFaults OverviewHardware Error Troubleshooting When a Fault Message is ShowingFault message Explanation Section Sensor ADC Error Sensor CPU ErrorSensor Incompatible Sensor RTD OpenGlass Z Too High Sensor RTD Out of RangeReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Readings drift Sensor can be calibrated, but the current is too lowProcess readings are erratic Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Simulating Inputs Chlorine Troubleshooting When no Error Message is Showing GeneralAnod CathSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage ASIA-PACIFIC Americas HeadquartersEurope GermanyWarranty Specifications subject to change without notice