MODEL FCL-1056

SECTION 8.0

 

MAINTENANCE

8.2CHLORINE SENSOR

8.2.1General.

When used in clean water, the chlorine sensor requires little maintenance. Generally, the sensor needs mainte- nance when the response becomes sluggish or noisy or when readings drift following calibration. For a sensor used in potable water, expect to clean the membrane every month and replace the membrane and electrolyte solu- tion every three months. In water containing large amounts of suspended solids, for example open recirculating cooling water, membrane cleaning or replacement will be more frequent.

8.2.2 Cleaning the membrane.

Clean the membrane with water sprayed from a wash bottle. Do not use tissues to clean the membrane.

8.2.3 Replacing the electrolyte solution and membrane.

CAUTION

Fill solution may cause irritation. May be harmful if swallowed. Read and follow manual.

1.Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 8-1.

2.Hold the sensor over a container with the cathode pointing down.

3.Remove the fill plug and allow the electrolyte solution to drain out.

4.Inspect the cathode. If it is tarnished, clean it using a cotton-tipped swab dipped in baking soda or alumina. Use type A dry powder alumina intended for metallographic polishing of medium and soft metals. Rinse thor- oughly with water.

5.Wrap the plug with two turns of pipe tape and set aside. Remove old tape first.

6.Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and mem- brane holder pointing up. Fill the cup with electrolyte solution and allow the wooden ring to soak up the solu- tion (usually takes several minutes).

7.Hold the sensor at about a 45-degree angle with the cathode end pointing up. Add electrolyte solution through the fill hole until the liquid overflows. Tap the sensor near the threads to release trapped air bubbles. Add more electrolyte solution if necessary.

8.Place the fill plug in the electrolyte port and begin screwing it in. After several threads have engaged, rotate the sensor so that the cathode is pointing up and continue tightening the fill plug. Do not overtighten.

9.Place a new O-ring in the groove around the cathode post. Cover the holes at the base of the cathode stem with several drops of electrolyte solution.

10.Insert a small blunt probe, like a toothpick with the end cut off, through the pressure equalizing port. See Figure 8-1.

NOTE

Do not use a sharp probe. It will puncture the bladder and destroy the sensor.

Gently press the probe against the bladder several times to force liquid through the holes at the base of the cathode stem. Keep pressing the bladder until no air bubbles can be seen leaving the holes. Be sure the holes remain covered with electrolyte solution.

11.Place a drop of electrolyte solution on the cathode, then place the membrane assembly over the cathode. Screw the membrane retainer in place.

12.The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte solution has been replenished.

54

Image 64
Contents FCL with 1056 Analyzer Essential Instructions English Free ChlorinePpm mg/L ORPLive/Continous S1 Free Cl PH Correction ManualS1 Manual pH Temp Units Calibrate Sensor 1 Free chlorine Menu TreeSensor 2 pH Alarms TemperatureDiagnostic Setup Reset AnalyzerAbout This Document Rev. Level DateTable of Contents FCL-1056Table of Contents CONT’D List of TablesList of Tables CONT’D IiiFeatures Section Description and SpecificationsApplications Specifications Sensor Specifications GeneralSpecifications Analyzer Accessories Ordering InformationComponent Parts PH Correction required selectionThis page left blank intentionally Section Installation Unpacking and InspectionFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstallation Installing the SensorsGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally Section Wiring POWER, ALARM, and Output WiringPower Analog output wiringSensor Wiring Alarm relay connectionsWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Section Display and Operation DisplayKeypad Programming the ANALYZER-TUTORIAL CalibrateOutputs RangeOutput Range O1 S1 4mA 0.000 ppm O1 S1 20mA 08.50 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityUsing Hold HoldConfiguring the Main Display Main FormatLanguage English Contrast Format, Language, Warning, and ContrastSection Programming the Analyzer Default SettingsGeneral Alarms Sensor assignmentChoices Default CONFIGURING, RANGING, and Simulating Outputs DefinitionsProgramOutput ConfigureAssign Configure Simulate Output Configure Procedure Ranging OutputsOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfigure Simulate Configuring Alarms and Assigning SetpointsRangeSimulate OutputSection Programming the Analyzer Configure/Setpoint AlarmsAlarm Alarms Configure/Setpoint Alarm Simulate AlarmSimulate Don’t SimulateSynch Timers Yes Alarms Configure/Setpoint SimulateProcedure Synchronizing Timers Definitions Chlorine Configuring the MeasurementDefinitions pH/ORP Measurement Sensor1SensorConfiguring Temperature Related Settings Definitions pHProgram Outputs Alarms Measurement TemperatureConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompProgram Alarms Measurement Temperature Setting UP DiagnosticsSecurity SecurityProgram Measurement Temperature Security Diagnostic SetupGI Fault High 1500MΩ Procedure Setting Up DiagnosticsResetting the Analyzer Program Temperature Security DiagnosticsReset Analyzer Reset AnalyzerSection Calibration Calibrating TemperatureIntroduction Procedure Calibrate SensorSensor11 OutputS1 Calibration S1 Calibration + 25.0CProcedure-Zeroing the Sensor Calibration Free ChlorineCalibrate SensorSensor11 Output S1 Calibration Choose Free ChlorineTemperature S1 Calibration ZeroZeroCalCal Process Cal Calibrate?Sensor Sensor OutputCalibration pH Calibration Slope and OffsetZeroBufferCalCal Auto10.01 pH 07.01 pHStable Time Manual BufferManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer Redox Temperature S2 pH Cal Buffer Cal StandardizeStandardizeSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset Output Calibration Analog OutputsMA Output Cal Meter 000 mA Trim Complete This page left blank intentionally Section Digital Communications Model FCL-1056 Section Section Maintenance AnalyzerCleaning the membrane Replacing the electrolyte solution and membraneChlorine Sensor GeneralOther Maintenance Cleaning the SensorPH Sensor Cleaning the flow controller Constant Head Flow ControllerWeight Weight Section Troubleshooting Using the Diagnostic FeatureFaults OverviewFault message Explanation Section Troubleshooting When a Fault Message is ShowingHardware Error Sensor CPU Error Sensor ADC ErrorSensor Incompatible Sensor RTD OpenSensor RTD Out of Range Glass Z Too HighReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Process readings are erratic Sensor can be calibrated, but the current is too lowReadings drift Sensor does not respond to changes in chlorine level Chlorine readings are too lowTroubleshooting When no Error Message is Showing pH Calibration Error During Two-Point CalibrationCalibration Error during Standardization Sensor Does Not Respond to Known pH ChangesPH Readings Are Moderately Noisy and Tend to Wander Troubleshooting When no Error Message is Showing General Simulating Inputs ChlorineAnod CathSimulating Inputs pH Simulating pH inputSimulating Inputs Temperature Simulating temperaturePage Americas Headquarters ASIA-PACIFICEurope GermanyWarranty Specifications subject to change without notice