Emerson PN 51-FCL-1056 instruction manual PH Sensor, Cleaning the Sensor, Other Maintenance

Page 65

MODEL FCL-1056

SECTION 8.0

 

MAINTENANCE

 

 

 

 

 

 

 

 

FIGURE 8-1. Chlorine Sensor Parts

 

 

 

 

 

 

 

 

 

TABLE 8-1. Spare Parts

 

 

 

 

 

 

33523-00

Electrolyte Fill Plug

 

9550094

O-Ring, Viton 2-014

 

33521-00

Membrane Retainer

 

23501-08

Free Chlorine Membrane Assembly: includes one membrane assembly and one O-ring

 

23502-08

Free Chlorine Membrane Kit: includes 3 membrane assemblies and 3 O-rings

9210356

#4 Free Chlorine Sensor Fill Solution, 4 oz (120 mL)

8.3 pH SENSOR

8.3.1 General.

When used in clean water, the pH sensor requires little maintenance. Generally, the sensor needs maintenance when the response becomes sluggish or noisy. In clean water the typical cleaning frequency is once a month. In water containing large amounts of suspended solids, for example open recirculating cooling water, cleaning fre- quency will be substantially greater.

8.3.2 Cleaning the Sensor

Remove soft deposits by rinsing with a stream of water from a wash bottle. If the sensor becomes coated with rust, dissolve the rust by soaking the sensor in dilute citric acid (dissolve 5 grams of citric acid crystals in 100 mL of water) for no longer than 30 minutes at room temperature. Rinse the sensor thoroughly with water and soak in pH 4 buffer for several hours. Recalibrate the sensor in buffers before returning it to service.

8.3.3 Other Maintenance

The 3900VP-02-10 pH sensor supplied with the Model FCL-02 is disposable. It has no replaceable parts.

55

Image 65
Contents FCL with 1056 Analyzer Essential Instructions Free Chlorine EnglishPpm mg/L ORPS1 Manual pH Temp Units S1 Free Cl PH Correction ManualLive/Continous Sensor 2 pH Menu TreeCalibrate Sensor 1 Free chlorine Temperature AlarmsDiagnostic Setup Reset AnalyzerRev. Level Date About This DocumentFCL-1056 Table of ContentsList of Tables Table of Contents CONT’DIii List of Tables CONT’DApplications Section Description and SpecificationsFeatures Specifications Analyzer Specifications GeneralSpecifications Sensor Ordering Information AccessoriesComponent Parts PH Correction required selectionThis page left blank intentionally Unpacking and Inspection Section InstallationFCL-01 free chlorine without continuous pH correction FCL-02 free chlorine with continuous pH correctionInstalling the Sensors InstallationGeneral Information Mounting, Inlet, and Drain ConnectionsModel FCL-01 This page left blank intentionally POWER, ALARM, and Output Wiring Section WiringPower Analog output wiringAlarm relay connections Sensor WiringWiring Diagram for Free Chlorine Sensor Model FCL-1056 Section Wiring Display Section Display and OperationKeypad Calibrate Programming the ANALYZER-TUTORIALOutputs RangeO1 S1 20mA 08.50 ppm Output Range O1 S1 4mA 0.000 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityHold Using HoldMain Format Configuring the Main DisplayLanguage English Contrast Format, Language, Warning, and ContrastGeneral Default SettingsSection Programming the Analyzer Sensor assignment AlarmsChoices Default Definitions CONFIGURING, RANGING, and Simulating OutputsAssign ConfigureProgramOutput Procedure Ranging Outputs Configure Simulate Output ConfigureOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CConfiguring Alarms and Assigning Setpoints Configure SimulateRangeSimulate OutputSection Programming the Analyzer Alarm AlarmsConfigure/Setpoint Alarm Simulate Alarm Alarms Configure/SetpointSimulate Don’t SimulateProcedure Synchronizing Timers Alarms Configure/Setpoint SimulateSynch Timers Yes Definitions pH/ORP Configuring the MeasurementDefinitions Chlorine Sensor1Sensor MeasurementDefinitions pH Configuring Temperature Related SettingsTemperature Program Outputs Alarms MeasurementConfiguring Security Settings S1 Temp Comp Auto S2 Temp CompSetting UP Diagnostics Program Alarms Measurement TemperatureSecurity SecurityDiagnostic Setup Program Measurement Temperature SecurityGI Fault High 1500MΩ Procedure Setting Up DiagnosticsProgram Temperature Security Diagnostics Resetting the AnalyzerReset Analyzer Reset AnalyzerIntroduction Calibrating TemperatureSection Calibration Calibrate SensorSensor11 Output ProcedureS1 Calibration S1 Calibration + 25.0CCalibration Free Chlorine Procedure-Zeroing the SensorChoose Free Chlorine Calibrate SensorSensor11 Output S1 CalibrationCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalSensor Sensor OutputCalibration Slope and Offset Calibration pHAuto ZeroBufferCalCalStable Time 07.01 pH10.01 pH Buffer ManualManual Buffer 1 0 7.00 pH S2 pH Manual Cal Buffer StandardizeStandardize Redox Temperature S2 pH Cal Buffer CalSlope 56.19 mV/pH Offset S1 Enter Value 00 pHProcedure-Entering a Known Slope and Offset MA Output Cal Meter 000 mA Trim Complete Calibration Analog OutputsOutput This page left blank intentionally Section Digital Communications Model FCL-1056 Section Analyzer Section MaintenanceReplacing the electrolyte solution and membrane Cleaning the membraneChlorine Sensor GeneralPH Sensor Cleaning the SensorOther Maintenance Constant Head Flow Controller Cleaning the flow controllerWeight Weight Using the Diagnostic Feature Section TroubleshootingFaults OverviewHardware Error Troubleshooting When a Fault Message is ShowingFault message Explanation Section Sensor ADC Error Sensor CPU ErrorSensor Incompatible Sensor RTD OpenGlass Z Too High Sensor RTD Out of RangeReference Impedance Too High Broken GlassTroubleshooting When a Warning Message is Showing Troubleshooting When no Error Message is Showing Chlorine Readings drift Sensor can be calibrated, but the current is too lowProcess readings are erratic Chlorine readings are too low Sensor does not respond to changes in chlorine levelCalibration Error During Two-Point Calibration Troubleshooting When no Error Message is Showing pHSensor Does Not Respond to Known pH Changes Calibration Error during StandardizationPH Readings Are Moderately Noisy and Tend to Wander Simulating Inputs Chlorine Troubleshooting When no Error Message is Showing GeneralAnod CathSimulating pH input Simulating Inputs pHSimulating temperature Simulating Inputs TemperaturePage ASIA-PACIFIC Americas HeadquartersEurope GermanyWarranty Specifications subject to change without notice